4.4.2 Gas calibration
Select the "Gas Calibration" in the menu, then press the middle key to enter it, input the
password (default password: 1111), then left- click to select the password cell, right-click to select
the password number, as shown in Figure 12.13.
After input the correct password, select the gas type, as shown in Figure 14. Press the left button
to return to the previous menu, press the right button to select the gas type, press the middle
button to enter the next menu, make the calibration to the selected gas, as shown in Figure 15.
As shown in Figure 15, right-click to select the calibration point, the calibration point values are
arranged in ascending order. Note: the value of the first calibration point should be less than the
value of the second calibration point, and the value of the second calibration point should be
less than the value of the third calibration point. When make a single point calibration, only the
first calibration point, no nee to make the the rest of the calibration points. Press the middle
button, as shown in Figure 16, press the left button to select, press the right button to increase
the value,
Press the middle button to confirm. After inputting the calibration point value, press the enter key,
the detector will calibrate itself,
It displays that the calibration is successful or failed, as shown in Figure 17.18.
21