Page 9 of 51

The system will default to the last window settings and automatically display them at next power on. 

1.4. Keypad Functions

This keypad is dual function keypad: 
1. When separately pressed, is shortcut function, referring to "2. Quickly set menu instructions ";

2. Press

and  Number  key,  is  Menu  key,  referring  to"6.Menu  Window  Description".Follow  these 

guidelines when using the Flowmeter keypad (Refer to Keypad Figure): 

  And 

To input numbers. 

Backspace or delete characters to the left. 

  And 

  Return  to  the  last  menu  or  to 

open  the  next  menu.  Acts  as  "+"  and  "-"  functions 

when entering numbers. 

  Select a menu. Press this key first, input two 

menu numbers and then enter the selected menu. For 

example  ,to  input  a  pipe  outside  diameter,  press 

keys, where "11" is the window 

ID to display the parameter for pipe outside diameter. 

1.5. Keypad Operation 

The flow meter adopts the window software design to consolidate or subdivide all of the parameters entered, the 

instrument setup and measurement result displays into more than 100 independent windows. The operator can 

input parameters, modify settings or display measurement results by "visiting" a specific window. These windows 

are arranged by 2-digit serial numbers (including "+" sign) from 00~99, then to +0, +1, etc. Each window serial 

number, or so-called window ID code, has a defined meaning. For example, Window M11 indicates the parameter 

input for pipe outside diameter, while Window M25 indicates the mounting spacing between the transducers, etc. 

(Refer

Windows Display Explanations). 

The  keypad  shortcut  to  visit  a  specific  window  is  to  press  the 

key  at  any  time,  then  input  the  2-digit 

window ID code. For example, to input or check the pipe outside diameter, just press the 

keys for window ID code 11. 

Another  method  to  visit  a  particular  window  is  to  press 

and 

keys  to  scroll  the  screen.  For 

example,  if  the  current  window  ID  code  is  M02,  press 

key  to  enter  Window  M01,  press  the 

button again to enter Window M00; then, press the 

  key to back Window M01, and press the 

  key 

again to enter Window M02. 
Windows are separated into three types: (1) Data Type, such as M11, M12; (2) Option Type, such as M14; (3) 

Pure Display Type, such as M01, M00. 
You  can  check  the  corresponding  parameters  by  visiting  the  Data  Type  Windows.  If  you  want  to  modify  the 
parameters, input the digits and press 

or press 

first, input the digits then press 

again to 

confirm.

 

UFM-40

 

 

  

Ultrasonic transit-time flowmeter

 

Содержание UFM-40

Страница 1: ......

Страница 2: ......

Страница 3: ...e spare parts are in accordance with the packing list Make sure that there is no damage to the enclosure due to a loose screw or loose wire or other damage that may have occurred during transportation Any questions please contact your representative as soon as possible UFM 40 Ultrasonic transit time flowmeter ...

Страница 4: ...ement Site Selection 14 4 Transducer Installation 15 4 1 Installing the transducers 15 4 1 1Transducer Spacing 15 4 1 2 Transducer Mounting Methods 15 4 1 3 V Method 15 4 1 4 Z Method 16 4 1 5 N Method not commonly used 16 4 2 Transducer Mounting Inspection 16 4 2 1 Signal Strength 16 4 2 2 Signal Quality Q value 17 4 2 3 Total Time and Delta Time 17 4 2 4 Transit Time Ratio 17 4 2 5 Warnings 17 5...

Страница 5: ......

Страница 6: ......

Страница 7: ......

Страница 8: ......

Страница 9: ...asurement results by visiting a specific window These windows are arranged by 2 digit serial numbers including sign from 00 99 then to 0 1 etc Each window serial number or so called window ID code has a defined meaning For example Window M11 indicates the parameter input for pipe outside diameter while Window M25 indicates the mounting spacing between the transducers etc Refer Windows Display Expl...

Страница 10: ......

Страница 11: ......

Страница 12: ......

Страница 13: ......

Страница 14: ...iquid such as a vertical pipe with flow in the upward direction or a full horizontal pipe Ensure enough straight pipe length at least equal to the figure shown below for the upstream and downstream transducers installation Ensure that the pipe surface temperature at the measuring point is within the transducer temperature limits Consider the inside condition of the pipe carefully If possible selec...

Страница 15: ......

Страница 16: ......

Страница 17: ......

Страница 18: ......

Страница 19: ......

Страница 20: ......

Страница 21: ......

Страница 22: ...eset Zero 44 Manual Zero Point 45 Scale Factor 46 Network identifying address code 47 System Lock Input and output setup 55 CL Mode Select 56 CL 4mA Output Value 57 CL 20mA Output Value 58 CL Check 59 CL Current Output 60 Date and Time 61 ESN 62 Serial Port Parameter 67 FO Frequency Range 68 Low FO Flow Rate 69 High FO Flow Rate 70 LCD Backlit Option 72 Working Timer 77 Beeper Setup 78 OCT Output ...

Страница 23: ...in Window M31 If the NEG Total has been turned off the NEG Total value displayed is the total prior to its turn off Date Time Flow Rate The time setting method can be found in WindowM60 03 04 03 15 49 40 R Flow 0 116 m3 h Heat Capacity Totalizer Heat Capacity Display Instantaneous Heat Capacity and Totalizer Heat Capacity Net Energy Totalizer E T Instantaneous Energy EFR Note when the instrument i...

Страница 24: ...R System Normal Diagnoses Net Flow Today Display Net Flow Today POS Flow Today 09 0 458748 m3 Pipe Outer Diameter Enter the pipe outside diameter or enter the pipe circumference in Window M10 The pipe outside diameter must range from 10mm to 6000mm Note Enter either pipe outside diameter or pipe circumference Pipe Outer Diameter 50mm Enter the pipe wall thickness Pipe Wall Thickness 4 00 mm Pipe M...

Страница 25: ......

Страница 26: ...y falls below this value the measurement stops automatically If the flow meter is already able to stop measuring when the pipe is empty a value in the range of 30 to 40 should also be entered in this window to ensure no measurement when the pipe is empty Holding with PoorSig NO Measurement Units Options Select the measurement unit as follows 0 Metric 1 English Measurement Units In 0 Metric Flow Ra...

Страница 27: ...1000 7 x 10000 1E 4 Factory default factor is x1 Totalizer Multiplier 0 x0 001 1E 3 ON OFF POS Totalizer On off POS Totalizer NO indicates the Flowmeter starts to totalize the value When it is turned off the positive totalizer displays in Window M02 will not change Factory default is YES POS Totalizer 35 YES ON OFF NEG Totalizer On off NEG Totalizer NO indicates the Flowmeter starts to totalize th...

Страница 28: ......

Страница 29: ......

Страница 30: ...lue that the user entered in Window M58 In the 4 20mA modes F4mAand F20mA can be selected as a positive or negative flow value as long as the two values are not the same CL 4mA Output Value CL 4 mA Output Value 0 0000 m3 h Set the CL output value according to the flow value at 4mA The flow unit s options are the same as those in Window m31 Once 4 20mA vs Vel is selected in Window M56 the unit shou...

Страница 31: ...ade in this window The format for setting time setting is 24 hours Press wait until appears the modification can be made ESN Display electronic serial number ESN of the instrument This ESN is the only one assigned to each Flowmeter ready to leave the factory The factory uses it for files setup and the user uses it for management Ultrasonic Flowmeter S N V2500008 Serial Port Settings This window is...

Страница 32: ......

Страница 33: ... the or to review totalizer in days months and years 0 for day 1 for month 2 for year Use the or to review the totalizer in some day some month some year For example to display the flow total for July 18 2000 the display at the upper right corner of the screen indicates that it was working properly the whole day On the contrary if G is displayed it indicates that the instrument gain was adjusted a...

Страница 34: ... are available day per day hour per hour min per minute sec per second The factory default unit is hour Temperature sensitivity and user SHC When the delta temperature is less than the sensitivity set energy will not be accumulated Set the adjustable temperature range of 0 to 10 The factory default setting is 0 2 When the user specific heat is setting energy is accumulated according to the user sp...

Страница 35: ......

Страница 36: ......

Страница 37: ......

Страница 38: ......

Страница 39: ......

Страница 40: ......

Страница 41: ...r lf 注0 Return daily instantaneous flow d ddddddE dd cr 1 DQH cr lf Return hourly instantaneous flow d ddddddE dd cr DQM cr lf Return instantaneous flow per minute d ddddddE dd cr DQS cr lf Return instantaneous flow per second d ddddddE dd cr DV cr lf Return instantaneous velocity d ddddddE dd cr DI cr lf Return positive accumulative flow dddddddE d cr 2 DI cr lf Return negative accumulative flow ...

Страница 42: ...ed Corresponding binary code is 57H 31H 32H 33H 34H 35H 44H 56H and 0DH Function sign Function sign can add up to 5 basic commands Prefix P is allowed together to form a compound command sent to the flowmeter together The flowmeter will answer simultaneously For example if No 4321 flowmeter is requested to simultaneously return 1 instantaneous flow 2 instantaneous flow velocity 3 positive total fl...

Страница 43: ...r Address Function Code Register Address Register Number CRC Verify Code Flowmeter returned data is assuming the current flow 1 234567m3 h 0x01 0x03 0x04 0x06 0x51 0x3F 0x9E 0x3B 0x32 Flowmeter Address Function Code Data Bytes Data 1 2345678 CRCVerify Code The four bytes 3F 9E 06 51 is in the IEEE754 format single precision floating point form of 1 2345678 Pay attention to the data storage order o...

Страница 44: ...B 40028 4 20mA Analog output low word 32 bits real 2 Unit mA 001C 40029 4 20mA Analog output high word 001D 40030 Error code char 1 2 String 3 Refer to Error Analysis for detailed codes meanings 001E 40031 Error code char 3 4 001F 40032 Error code char 5 6 003B 40060 Velocity unit char 1 2 String 2 Currently support m s only 003C 40061 Velocity unit char 3 4 003D 40062 Flow unit char 1 2 String 2 ...

Страница 45: ...ut 4 Ball value 5 Mounting base 6 Sensor 10 2 Measurement Point Selection To guarantee highly accurate measurement results it is necessary to select an appropriate measurement point before installing the transducer For examples of measuring point selection see the related section in the manual 10 3 Determining Transducer Spacing Transducer Installation The mounting space of insertion transducer is...

Страница 46: ......

Страница 47: ......

Страница 48: ...nsducer mounting method and press the key to confirm Choose according to the pipes on site Transducer Mounting 1 Z Step6 Adjust Transducer spacing Press the keys to enter the window M25 accurately install the transducer according to the displayed transducer mounting spacing and the selected mounting method Refer to Installing the Transducers in this chapter Transducer Spacing 192 68 mm Step7 Displ...

Страница 49: ......

Страница 50: ......

Страница 51: ...ter temperature is 100 0 05 Temperature difference is 0 00 0 05 7 After the ice water mixture temperature sensors back to room temperature put them into a constant temperature oil water bath the temperature is 100 C 11 5 Installation of RTD Module Before installation of the RTD module as shown in the figure below After installation as shown in the figure below UFM 40 Ultrasonic transit time flowme...

Отзывы: