Page 44 of 51

$0000

40001 

Flow/s - low word

 

32 bits real 

$0001

40002 

Flow/s - high word

 

$0002

40003 

Flow/m - low word

 

32 bits real 

$0003

40004 

Flow/m- high word

 

$0004

40005 

Flow/h - low word

 

32 bits real 

$0005

40006 

Flow/h - high word

 

$0006

40007 

Velocity – low word

 

32 bits real 

$0007

40008 

Velocity – high word

 

$0008

40009 

Positive total – low word

 

32 bits uint. 

$0009

40010 

Positive total – high word

 

$000A 

40011 

Positive total – exponent

 

16 bits int. 

$000B 

40012 

Negative total—low word 

32 bits int. 

$000C 

40013 

Negative total—high word 

$000D 

40014 

Negative total—exponent 

16 bits int. 

$000E 

40015 

Net total—low word 

32 bits int. 

$000F

40016 

Net total

high word 

$0010

40017 

Net total—exponent 

16 bits int. 

$0016

40023 

Up signal int – low word

 

32 bits real 

0

99.9 

$0017

40024 

Up signal int – high word

 

$0018

40025 

Down signal int – low word

 

32 bits real 

0

99.9 

$0019

40026 

Down signal int – high word

 

$001A 

40027 

Quality

 

16 bits int. 

0

99 

$001B 

40028 

4~20mA

 Analog output

low 

word

 

32 bits real 

Unit: mA

 

$001C 

40029 

4~20mA

 Analog output

high 

word

 

$001D 

40030 

Error code – char 1,2

 

String 

Refer to "Error 

Analysis" for 

detailed codes 

meanings.

 

$001E 

40031 

Error code – char 3,4

 

$001F 

40032 

Error code – char 5,6

 

$003B 

40060 

Velocity unit – char 1,2

 

String

Currently support 

m/s only 

$003C 

40061 

Velocity unit – char 3,4

 

$003D 

40062 

Flow unit – char 1,2

 

String

2

$003E 

40063 

Flow unit – char 3,4

 

$003F

40064 

Total unit – char 1,2

 

String

1

Notes: 

1. The following flow rate units are available:

0. "m3"

   

Cubic Meter 

1. "l"

Liters 

2. "ga"

   

Gallons 

3. "ig"

   

Imperial Gallons 

4. "mg"

   

Million Gallons 

5. "cf"

   

Cubic Feet 

6. "ba"

   

US Barrels 

7. "ib"

   

Imperial Barrels 

8. "ob"

Oil Barrels 

2. When the flowmeter address or communication baud rate change, the meter will work under the new

address or communication baud rate after the communication baud rate responded with returned primary

address and communication baud rate.

3. 16  bits  int

short  integer,  32  bits  int

long  integer,  32  bits  real

floating  point  number,  String

alphabetic string.

 

UFM-40

 

 

  

Ultrasonic transit-time flowmeter

 

PDU 

Address

 

Register

 

Read

 

Write

 

Type

 

No. registers*

 

Содержание UFM-40

Страница 1: ......

Страница 2: ......

Страница 3: ...e spare parts are in accordance with the packing list Make sure that there is no damage to the enclosure due to a loose screw or loose wire or other damage that may have occurred during transportation Any questions please contact your representative as soon as possible UFM 40 Ultrasonic transit time flowmeter ...

Страница 4: ...ement Site Selection 14 4 Transducer Installation 15 4 1 Installing the transducers 15 4 1 1Transducer Spacing 15 4 1 2 Transducer Mounting Methods 15 4 1 3 V Method 15 4 1 4 Z Method 16 4 1 5 N Method not commonly used 16 4 2 Transducer Mounting Inspection 16 4 2 1 Signal Strength 16 4 2 2 Signal Quality Q value 17 4 2 3 Total Time and Delta Time 17 4 2 4 Transit Time Ratio 17 4 2 5 Warnings 17 5...

Страница 5: ......

Страница 6: ......

Страница 7: ......

Страница 8: ......

Страница 9: ...asurement results by visiting a specific window These windows are arranged by 2 digit serial numbers including sign from 00 99 then to 0 1 etc Each window serial number or so called window ID code has a defined meaning For example Window M11 indicates the parameter input for pipe outside diameter while Window M25 indicates the mounting spacing between the transducers etc Refer Windows Display Expl...

Страница 10: ......

Страница 11: ......

Страница 12: ......

Страница 13: ......

Страница 14: ...iquid such as a vertical pipe with flow in the upward direction or a full horizontal pipe Ensure enough straight pipe length at least equal to the figure shown below for the upstream and downstream transducers installation Ensure that the pipe surface temperature at the measuring point is within the transducer temperature limits Consider the inside condition of the pipe carefully If possible selec...

Страница 15: ......

Страница 16: ......

Страница 17: ......

Страница 18: ......

Страница 19: ......

Страница 20: ......

Страница 21: ......

Страница 22: ...eset Zero 44 Manual Zero Point 45 Scale Factor 46 Network identifying address code 47 System Lock Input and output setup 55 CL Mode Select 56 CL 4mA Output Value 57 CL 20mA Output Value 58 CL Check 59 CL Current Output 60 Date and Time 61 ESN 62 Serial Port Parameter 67 FO Frequency Range 68 Low FO Flow Rate 69 High FO Flow Rate 70 LCD Backlit Option 72 Working Timer 77 Beeper Setup 78 OCT Output ...

Страница 23: ...in Window M31 If the NEG Total has been turned off the NEG Total value displayed is the total prior to its turn off Date Time Flow Rate The time setting method can be found in WindowM60 03 04 03 15 49 40 R Flow 0 116 m3 h Heat Capacity Totalizer Heat Capacity Display Instantaneous Heat Capacity and Totalizer Heat Capacity Net Energy Totalizer E T Instantaneous Energy EFR Note when the instrument i...

Страница 24: ...R System Normal Diagnoses Net Flow Today Display Net Flow Today POS Flow Today 09 0 458748 m3 Pipe Outer Diameter Enter the pipe outside diameter or enter the pipe circumference in Window M10 The pipe outside diameter must range from 10mm to 6000mm Note Enter either pipe outside diameter or pipe circumference Pipe Outer Diameter 50mm Enter the pipe wall thickness Pipe Wall Thickness 4 00 mm Pipe M...

Страница 25: ......

Страница 26: ...y falls below this value the measurement stops automatically If the flow meter is already able to stop measuring when the pipe is empty a value in the range of 30 to 40 should also be entered in this window to ensure no measurement when the pipe is empty Holding with PoorSig NO Measurement Units Options Select the measurement unit as follows 0 Metric 1 English Measurement Units In 0 Metric Flow Ra...

Страница 27: ...1000 7 x 10000 1E 4 Factory default factor is x1 Totalizer Multiplier 0 x0 001 1E 3 ON OFF POS Totalizer On off POS Totalizer NO indicates the Flowmeter starts to totalize the value When it is turned off the positive totalizer displays in Window M02 will not change Factory default is YES POS Totalizer 35 YES ON OFF NEG Totalizer On off NEG Totalizer NO indicates the Flowmeter starts to totalize th...

Страница 28: ......

Страница 29: ......

Страница 30: ...lue that the user entered in Window M58 In the 4 20mA modes F4mAand F20mA can be selected as a positive or negative flow value as long as the two values are not the same CL 4mA Output Value CL 4 mA Output Value 0 0000 m3 h Set the CL output value according to the flow value at 4mA The flow unit s options are the same as those in Window m31 Once 4 20mA vs Vel is selected in Window M56 the unit shou...

Страница 31: ...ade in this window The format for setting time setting is 24 hours Press wait until appears the modification can be made ESN Display electronic serial number ESN of the instrument This ESN is the only one assigned to each Flowmeter ready to leave the factory The factory uses it for files setup and the user uses it for management Ultrasonic Flowmeter S N V2500008 Serial Port Settings This window is...

Страница 32: ......

Страница 33: ... the or to review totalizer in days months and years 0 for day 1 for month 2 for year Use the or to review the totalizer in some day some month some year For example to display the flow total for July 18 2000 the display at the upper right corner of the screen indicates that it was working properly the whole day On the contrary if G is displayed it indicates that the instrument gain was adjusted a...

Страница 34: ... are available day per day hour per hour min per minute sec per second The factory default unit is hour Temperature sensitivity and user SHC When the delta temperature is less than the sensitivity set energy will not be accumulated Set the adjustable temperature range of 0 to 10 The factory default setting is 0 2 When the user specific heat is setting energy is accumulated according to the user sp...

Страница 35: ......

Страница 36: ......

Страница 37: ......

Страница 38: ......

Страница 39: ......

Страница 40: ......

Страница 41: ...r lf 注0 Return daily instantaneous flow d ddddddE dd cr 1 DQH cr lf Return hourly instantaneous flow d ddddddE dd cr DQM cr lf Return instantaneous flow per minute d ddddddE dd cr DQS cr lf Return instantaneous flow per second d ddddddE dd cr DV cr lf Return instantaneous velocity d ddddddE dd cr DI cr lf Return positive accumulative flow dddddddE d cr 2 DI cr lf Return negative accumulative flow ...

Страница 42: ...ed Corresponding binary code is 57H 31H 32H 33H 34H 35H 44H 56H and 0DH Function sign Function sign can add up to 5 basic commands Prefix P is allowed together to form a compound command sent to the flowmeter together The flowmeter will answer simultaneously For example if No 4321 flowmeter is requested to simultaneously return 1 instantaneous flow 2 instantaneous flow velocity 3 positive total fl...

Страница 43: ...r Address Function Code Register Address Register Number CRC Verify Code Flowmeter returned data is assuming the current flow 1 234567m3 h 0x01 0x03 0x04 0x06 0x51 0x3F 0x9E 0x3B 0x32 Flowmeter Address Function Code Data Bytes Data 1 2345678 CRCVerify Code The four bytes 3F 9E 06 51 is in the IEEE754 format single precision floating point form of 1 2345678 Pay attention to the data storage order o...

Страница 44: ...B 40028 4 20mA Analog output low word 32 bits real 2 Unit mA 001C 40029 4 20mA Analog output high word 001D 40030 Error code char 1 2 String 3 Refer to Error Analysis for detailed codes meanings 001E 40031 Error code char 3 4 001F 40032 Error code char 5 6 003B 40060 Velocity unit char 1 2 String 2 Currently support m s only 003C 40061 Velocity unit char 3 4 003D 40062 Flow unit char 1 2 String 2 ...

Страница 45: ...ut 4 Ball value 5 Mounting base 6 Sensor 10 2 Measurement Point Selection To guarantee highly accurate measurement results it is necessary to select an appropriate measurement point before installing the transducer For examples of measuring point selection see the related section in the manual 10 3 Determining Transducer Spacing Transducer Installation The mounting space of insertion transducer is...

Страница 46: ......

Страница 47: ......

Страница 48: ...nsducer mounting method and press the key to confirm Choose according to the pipes on site Transducer Mounting 1 Z Step6 Adjust Transducer spacing Press the keys to enter the window M25 accurately install the transducer according to the displayed transducer mounting spacing and the selected mounting method Refer to Installing the Transducers in this chapter Transducer Spacing 192 68 mm Step7 Displ...

Страница 49: ......

Страница 50: ......

Страница 51: ...ter temperature is 100 0 05 Temperature difference is 0 00 0 05 7 After the ice water mixture temperature sensors back to room temperature put them into a constant temperature oil water bath the temperature is 100 C 11 5 Installation of RTD Module Before installation of the RTD module as shown in the figure below After installation as shown in the figure below UFM 40 Ultrasonic transit time flowme...

Отзывы: