14
IRT807
OFF
ZERO
kHz
Temp
SMOOTH
mV
mA
INSULA
TION
V
V
50V
1000V
INSULATION TESTER
Apo
MIN/MAX
LOCK
Hz
Test Voltage
F C
SELECT
TEST
RANGE
PI/DAR
A-HOLD
0
0.1 M
1 M
10 M
100 M
1 G
2 G
V
DC
M
IRT807
Apo
m
A
• Default =
mA
• Press x1 =
mA
SELECT
mAmps AC/DC
WARNING
Never connect the test leads across
(in parallel with) a voltage source
when attempting to measure current.
This can cause damage to the circuit
under test or this meter.
Note:
To measure current, you must
create a break in the circuit under test
and make the meter part of the circuit.
Two connection points are created when a
circuit is broken. On one side is the power
source and the other is the load.
Features:
SELECT
MIN/MAX
Test Voltage
F C
RANGE
LOCK
A-HOLD
Hz
PI/DAR
AC and DC CURRENT
RANGE
RESOLUTION
ACCURACY
AC (45Hz to 1000Hz)
40mA
0.01mA
±(1.5% +5 dgts)
1
400mA
0.1mA
DC
40mA
0.01mA
±(0.5% +5 dgts)
400mA
0.1mA
1. 1kHz bandwidth
Overload Protection 440mA, 1000V, FAST fuse
IRT807
OFF
ZERO
kHz
Temp
SMOOTH
mV
mA
INSULA
TION
V
V
50V
1000V
INSULATION TESTER
Apo
MIN/MAX
LOCK
Hz
Test Voltage
F C
SELECT
TEST
RANGE
PI/DAR
A-HOLD
0
0.1 M
1 M
10 M
100 M
1 G
2 G
V
DC
M
IRT807
Apo
m
A
Frequency
RANGE
RESOLUTION
ACCURACY
99.99Hz
0.01Hz
±(0.1% +3 dgts)
999.9Hz
0.1Hz
9.999kHz
0.001kHz
99.99kHz
0.01kHz
Sensitivity: >AC 1V RMS, >AC 10mA
The meter measures the frequency of
a voltage or current signal by counting
the number of times the signal crosses a
threshold level each second. To measure
Frequency, set selector dial Volt or Milliamp
position. In the mA position, press the
SELECT button
to select AC, if needed.
Press the
PI/DAR button
.
PI/DAR
• Default =
mA
Features:
LOCK
A-HOLD
Hz
PI/DAR
MIN/MAX
Test Voltage
F C
RANGE
SELECT