Auto On/APO
• Press any button to power on meter.
• Press and hold
AC/DC Volts/Power
button to power off meter.
• Auto power off (Apo) after 10 minutes of inactivity.
• Press and hold the Hold button when meter is in Hold mode to disable Auto
power off mode. Press and hold again to enable Auto power off mode.
• Default =
• Press x1 =
• Press x2 =
• Default =
• Press x1 =
• Press x2 =
CAT III
600V
VΩ
DL200
600V CAT III 200A
HOLD
HOLD
AC/DC Voltage: <600V AC/DC
WARNING
•
Do not attempt to measure more than 600V AC/DC.
•
Keep hands below line when measuring high current levels.
• Do not exceed 600 volts AC or DC – RMS on either the common or
multifunction test leads as measured from earth ground.
Select AC or DC Voltage.
Features:
HOLD
AC Volts
Ranges
Resolution
Accuracy
Overload Protection
1V to 199.9V
0.1V
±(1.5% + 5 dgts)
600V
200V to 600V
1V
DC Volts
Ranges
Resolution
Accuracy
Overload Protection
1V to 19.9V
0.01V
±(1.5% + 3 dgts)
600V
20V to 199.9V
0.1V
200V to 600V
1V
Minimum measurement AC mV: 500mV/DC mV:40mV
Response time (input; AC 100V): 1 sec
Max displayed value: 660V AC/DC
High warning symbol input voltage > AC30V/ > DC30V
Meter beeps input voltage > AC 601V/ > DC 601V
Input impedance: 10MΩ
Frequency Range: sine wave: 45Hz to 400Hz
Overload protection: 600V AC RMS or 600V DC
Resistance: <2000Ω
Features:
HOLD
WARNING
• Do not measure resistance on a live circuit.
Resistance
Ranges
Resolution
Accuracy
Overload Protection
200Ω to 2000Ω
0.1Ω to 1Ω
±(0.8% + 4 dgts)
600V
Max displayed value: 2200Ω
Overload protection: 600V AC RMS or 600V DC