background image

C50/C75-MAN    

P. 7

Charging

Ensure that you use the correct charger. This unit uses a 9V DC 
regulated charger.

Ensure that the batteries are fitted in the correct manner, and then
charge for at least 16 hours. Subsequent charges should be overnight.
NiMH batteries may be charged at any time, even for short periods to
conduct testing.

WARNING!

Under 

NO

c i rcumstance should you expose batteries to ex t reme heat or

f i re as they may explode and cause injury. Always dispose of old batteries
p romptly in a manner consistent with local disposal re gu l a t i o n s .

Emptying and Cleaning the In-line Water Trap

The in-line water trap should be checked and emptied on a regular
basis. Water vapor will condense in the probe line, which may cause the
water trap to fill suddenly if the probe is moved. Care should be taken
at all times.

Carefully remove the rubber plug from the bottom of the water-trap
housing. Dispose of the condensate in a suitable drain, care must be
taken as it could be acidic.

If condensate spills onto the skin or clothing, clean off immediately
using fresh water, seek medical advice if problems occur. Ensure plug is
replaced before performing combustion tests.

Changing the Particle Filter

This is a very important part of the analyzer and should be changed 
regularly. It prevents dust and dirt particles from entering the pump 
and sensors that will cause damage. The filter MUST be changed when
it appears discolored.

Remove water-trap assembly from the analyzer as shown above.
Remove the filter and plastic holder from the housing. Discard the filter
element but keep the holder to fit to the new filter. Clean the inside of
the filter housing with a suitable soft cloth. Fit the holder onto the new
filter element and then insert into the housing. Refit the housing onto
the analyzer.

Hook slides

out of dock

Slide outward

Pull plug out 

to drain

Insert a new filter

M a i n t e n a n c e

Periodic Service

WARNING!

Repair and service of this instrument is to be performed by qualified
personnel only. Improper repair or service could result in physical
degradation of the instrument. This could alter the protection from 
personal injury this meter provides to the operator. Perform only those
maintenance tasks that you are qualified to do.

These guidelines will help you attain long and reliable service from 
your meter:

• Calibrate your instrument annually to ensure it meets original

performance specifications

• Keep your instrument dry. If it gets wet, wipe dry immediately.

Liquids can degrade electronic circuits

• Whenever practical, keep the instrument away from dust and

dirt that can cause premature wear

• Although your instrument is built to withstand the rigors of daily

use, it can be damaged by severe impacts. Use reasonable
caution when using and storing the meter

Cleaning

Periodically clean your instruments case using a damp cloth. 

DO NOT

use abrasive, flammable liquids, cleaning solvents, or strong detergents
as they may damage the finish, impair safety, or affect the reliability of
the structural components.

Battery Replacement

This meter has been designed for use with both alkaline and 
rechargeable Nickel Metal Hydride (NiMH) batteries. No other types 
are recommended. The analyzer is supplied with 4 “AA” size alkaline
batteries. These should be installed into the instrument as shown below
(Fig 1) and indicated on the back of the unit.

CAUTION!

Take great care when installing the batteries to observe correct polarity.
Always check the meter for operation immediately after installing 
new batteries.

Using Re-Chargeable Batteries

The battery charger must only be used when NiMH batteries are fitted.

Alkaline batteries are not re-Chargeable. Attempting to recharge alkaline
batteries may result in damage to the product and may create a fire risk.

(Fig 1)

Note: Battery Direction

Battery Compartment

Cover

Back of

Instrument

Содержание C50

Страница 1: ...C50 C75 INSTRUCTION MANUAL 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com ...

Страница 2: ...ON and OFF Hold to scroll through top line display also selectable through menu 6 Hold Store Freezes reading on display Entire display flashes Press and hold for 2 seconds to store data in memory 7 Rotary Selector 8 Particle Filter 9 Water Trap 10 Analyzer Connections Main Display Parameters O2 Oxygen reading in percentage Tf Temperature is measured by the flue gas probe in Centigrade or Fahrenhei...

Страница 3: ...is the ratio of measured CO divided by calculated CO2 It gives an indication of How good a gas sample the instrument is reading How clean the boiler is running For example A new or clean domestic boiler will display a ratio of less than 0 004 a unit in need of cleaning 0 004 0 008 and a unit in need of major overhaul will show greater than 0 008 This only shows a reading when a combustion test is ...

Страница 4: ... sampling point be located at least two flue diameters downstream of any bend as close to the source as possible and that the probe tip is in the center of the flue With balanced flues and other domestic units the probe should be positioned far enough into the flue so that no air can back flush into the probe The probe depth stop cone provided with the instrument allows the probe to be used in hol...

Страница 5: ...e above the function keys The three keys are increases decrease and enter Set Time 1 Press ENTER 2 Use the up and down keys to select the correct time 3 Press ENTER to move to the next digit Note Time is displayed in military format example 7 00 pm is 19 00 Date 1 Press ENTER 2 Select desired format for the date display MM DD YY for month day year YY MM DD for year month day DD MM YY for day month...

Страница 6: ...STORE view mode the rotary dial is used to select the saved parameter to be displayed Use the UP DOWN keys to change the test number To exit the view mode press ENTER Storing Test Results At any time during a test you may store the readings in one of the 1 memory positions Press and hold for 2 seconds or more Printing Information Supplied as an accessory for the analyzer is an infrared thermal pri...

Страница 7: ...r x 2488 2 1 Tf 4 2 Ti Simplified 9 x H2 H2O Qgr x 2425 x 1 0 001 Tnet Wet loss K3 1 0 001 x Tnet Where K3 9 x H2 H2O Qgr x 2425 Combustion Combustion Theory In its simplest form combustion is the combining of oxygen O2 from the air with hydrogen H and carbon C from the fuel to form carbon dioxide CO2 water H2O and energy light and heat Perfect combustion occurs when all of the carbon and hydrogen...

Страница 8: ...ir free In the example we would be breaking the limit and corrective action should be taken to reduce the level of CO Air free values prevent false readings being submitted e g allowing more air into the boiler will increase the oxygen level in the flue and dilute any toxic gas reading Air free referencing gives readings as if they were undiluted What Does Oxygen Reference Mean Oxygen referencing ...

Страница 9: ...k Slide outward Pull plug out to drain Insert a new filter Maintenance Periodic Service WARNING Repair and service of this instrument is to be performed by qualified personnel only Improper repair or service could result in physical degradation of the instrument This could alter the protection from personal injury this meter provides to the operator Perform only those maintenance tasks that you ar...

Страница 10: ... 10 V AC off load Annual Re Calibration While the sensor has an expected life of more than two years in normal use it is recommended that the analyzer is re calibrated at least annually This is so that long term drift on the sensor and electronics can be eliminated local regulations may require more frequent re calibration and users should check with appropriate authorities to ensure the comply wi...

Страница 11: ...ited to implied warranties of merchantability and fitness for a particular purpose are limited to the express warranty UEi shall not be liable for loss of use of the instrument or other incidental or consequential damages expenses or economic loss or for any claim or claims for such damage expenses or economic loss A purchase receipt or other proof of original purchase date will be required before...

Отзывы: