REPLACING THE BATTERIES

This meter has been designed for use with both Alkaline or rechargeable Nickel Metal Hydride (NiMH) batteries. No other types are 

recommended and will void warranty. The analyzer is supplied with 3 (AA) size NiMH rechargeable batteries. These should be installed 

into the instrument.

  

CAUTION

Take great care when installing the batteries to observe correct polarity.  Always check the meter 

for operation immediately after installing new batteries.

Using Rechargeable Batteries

The battery charger must only be used when NiMH batteries are fitted. Alkaline batteries are not rechargeable. Attempting to recharge 

alkaline batteries may result in damage to the product and create a fire risk.

When changing the batteries, the time and date will need setting.

  

WARNING

Do not try and charge the unit with Alkaline batteries fitted. Do not mix NiMH cells of different capacities from different  

 

manufacturers, all cells must be identical.

Turn over the analyzer, remove the protective rubber boot and fit 3 “AA” batteries in the battery compartment. 

Take great care to 

ensure they are fitted with the correct battery polarity. 

Replace the battery cover and the protective rubber boot.

Time and Date

When changing the batteries, the time and date will need setting.

Battery Disposal 

Always dispose of depleted batteries using approved disposal methods that protect the environment.

Battery Charging

Ensure that you use the correct charger. This unit uses a 5V regulated charger. Ensure the batteries are fitted in the correct manner, 

and charge for at least 16 hours. Subsequent charges should be overnight. NiMH batteries may be charged at any time, even for short 

periods to conduct testing.

  

WARNING

Under NO circumstances should you expose batteries to extreme heat or fire as they may explode and cause injury. Always dispose of 

old batteries promptly in a manner consistent with local disposal regulations.

ANNUAL RECERTIFICATION

What You Get

Calibration (or sensor replacement)
30-point inspection by factory trained professionals
Firmware and software updates
Certificate of calibration
Another 12-month warranty on the analyzer
Return unit freight paid – within continental US and Canada
List of work performed, and parts replaced

www.calcert.com

[email protected]

1.800.544.2843

0

5

10

15

20

25

30

Содержание C165-N

Страница 1: ...C o m T e s t Prs Tem p Exch Test Room CO Combustion Analyzer C165 CO C O 2 T E M P P R S INSTRUCTION MANUAL ENGLISH C165 Residential Commercial Combustion Analyzer www calcert com sales calcert com...

Страница 2: ...ane Wireless Printer App Home Android Screen 16 Kane Wireless Printer App Android Settings Screen 16 Kane Wireless Printer App Android Personal Details Screen 16 Temperature Testing 17 Pressure Temper...

Страница 3: ...zer must only be used in well ventilated locations by trained and competent persons after due consideration of all potential hazards Users of portable gas detectors are recommended to conduct a bump c...

Страница 4: ...ctive Rubber Boot with Magnets O Serial Number QR Code Serial Number viewable under Protective Boot P Sensors Fitted label under Protective Boot Sensors that can be fitted in unit when shipped CO H2 C...

Страница 5: ...Analyzer C165 CO C O 2 T E M P P R S ANALYZER OVERVIEW CONT A Sensors Fitted CO NO CO H2 CO2 O2 Serial No UEiTEST INSTRUMENTS B C D E F G H I J M N T U W V X L K Q S R P O T2 T1 P2 P1 TOP BOTTOM Narro...

Страница 6: ...llation of boilers A low flow detection system warns of low flow and switches the pump off This also helps to prevent water ingress from overfilled water traps The analyzer has a protective rubber boo...

Страница 7: ...r O2 prior to shutting the unit off The meter will not switch off until the CO reading is below 20 ppm CAUTION The probe will be hot from flue gases Remove the probe from the flue and allow it to cool...

Страница 8: ...e flue Use the flue probes depth stop cone to set the position With balanced flues make sure the probe is positioned into the flue so no air can back flush into the probe CAUTION Ensure the flue probe...

Страница 9: ...o EXIT the MENU at any time simply move the rotary switch to any position other than Menu Any changes that have not been entered will be ignored MAIN MENU SUB MENU OPTIONS COMMENTS UNITS FUEL TYPE NAT...

Страница 10: ...an Menu Any changes that have not been entered will be ignored MAIN MENU SUB MENU OPTIONS COMMENTS REPORT AUX Stored AUX tests VIEW DEL ALL BACK COMBUSTION Stored combustion tests VIEW DEL ALL BACK CO...

Страница 11: ...ystem conditions are stable The rotary switch can be used to display the following information FLUE 1 Fuel type can be changed via MENU or STATUS setting Carbon Dioxide Oxygen left after combustion Sh...

Страница 12: ...option SCREEN Press Use and to navigate to the sub menu option AUX Press The display will show Press and a third line will appear Use and to navigate to the desired parameter to be displayed on line...

Страница 13: ...ER INTEGRITY TEST There are many methods to test heat exchange integrity One of these is to observe the Excess Air O2 and CO readings both before and after the blower turns on If the heat exchanger is...

Страница 14: ...Use or Arrow Buttons to select Printer Press button Use or Arrow Buttons to select Printer type Wireless KM IRP Kane IRP 2 Serial Wireless Printer App Please see pages 15 and 16 To charge Printer if o...

Страница 15: ...T screen Select EMAIL PRINTOUT A message window will open Enter email address Add message notes to Printout Send message Select CLEAR CONSOLE to delete a Printout Press to Confirm deletion To print Sa...

Страница 16: ...nect an analyzer Select disconnect from Printout screen Adding Notes Emailing from the Kane Wireless Printer App After a report is on the app Press the Pen icon on the top right of the app screen Pres...

Страница 17: ...low Temp Eg Return Temp Differential Temp Press to print a full pressure test or send to Smart Device via optional Wireless module Hold for 2 seconds to log a pressure report PRESSURE MEASUREMENT GOOD...

Страница 18: ...5 reading unspecified 0 3 volume 1 0 reading 1 0 reading 0 2 reading 5 reading Nitric Oxide Sensor optional 0 to 100ppm Overrange to 1500ppm 1ppm 2ppm 30ppm 1 5ppm 100ppm 1 5 reading 100ppm Pressure d...

Страница 19: ...017 IEC62321 7 1 2015 IEC62321 6 2015 AHRI 1260 ELECTROMAGNETIC COMPATIBILITY European Council Directive 89 336 EEC requires electronic equipment not to generate electromagnetic disturbances exceeding...

Страница 20: ...O2 CO2 CO Stack Temp Efficiency Hi Low fire Gas Pressure Supply and Return Water temp 80 Furnace Verify proper combustion O2 CO2 CO Stack Temp Vent Pressure Efficiency Verify Set Up Gas Pressure Test...

Страница 21: ...ssure Temperature Rise AC side Static Pressure Drop across coils WHAT RESULTS ARE GENERALLY ACCEPTABLE Atmospheric Gas Fixed Burners Gas Fired Burners Oil Fired burners 2 Oil Fuel Positive Overfire Ga...

Страница 22: ...vice of this instrument is to be performed by qualified personnel only Improper repair or service could result in physical degradation of the instrument This could alter the protection from personal i...

Страница 23: ...tment Take great care to ensure they are fitted with the correct battery polarity Replace the battery cover and the protective rubber boot Time and Date When changing the batteries the time and date w...

Страница 24: ...ak detectors Any accessories that are returned will be checked Packing Your Analyzer Put analyzer and probes back in their cases The case should be put into a box with 1 2 inches of packing on each si...

Страница 25: ...temperature and then reference a slide scale that would give you the relative efficiency number UEi s electronic combustion analyzers perform the measurements on a continuous basis and can calculate t...

Страница 26: ...e inoperative from such defects the unit will be repaired or replaced at UEi s option This warranty covers normal use and does not cover damage which occurs in shipment or failure which results from a...

Отзывы: