PRODUCT SPECIFICATION FOR HANDHELD AUTO4-2 AND AGA5000
*1
Using dry gases at STP
*_ Standard on models AGA5000 only
To obtain the quoted specification an instrument should be calibrated with clean ambient air (normally outside the workshop) at standard
temperature and pressure (STP).
Note: The analyzer is not for use with a Dynometer, unless the high temperature probe is fitted.
Parameter
Resolution
Accuracy
Range
Carbon Monoxide
0.01 %
+/- 5 % of reading
*1
0-10 %
(Infrared)
+/- 0.5 % volume
*1
Over-range: 20 %
Oxygen
0.01%
+/- 5 % of reading
*1
0-21%
(fuel cell)
+/- 0.1 % of reading
*1
Over-range: 48%
Hydrocarbon
1 ppm
+/- 5 % of reading
*1
0-5000 ppm
(Infrared)
+/- 12 ppm volume
*1
Over-range: 10,000 ppm
Carbon Dioxide
0.1%
+/- 5% of reading
*1
0-16%
(Infrared)
+/- 0.5% ppm volume
*1
Over-range: 25%
Nitric Oxide*_
1 ppm
0-4000 ppm +/-4% or
0-5000 ppm
(fuel cell)
25 ppm; 4000-5000 ppm
+/-5%
Oil Temperature
1.0˚C/F
±2.0˚C ±0.3% of reading
0-150˚C
±3.6˚F ±0.3% of reading
32-302˚F
RPM
1 rpm
50 rpm
200-6,000 rpm
Carbon Monoxide
0.01%
Calculated
0-15%
Corrected CO
Lambda
0.001
0.8 - 1.2
AFR (Gasoline)
00.01
11.76 - 17.64
(LPG)
00.01
12.48 - 18.72
Sensor response T
95
Nominal 20 seconds Auto4-2 and AGA5000
Warm Up
Less than 3 minutes
Pre-programmed Fuels
Gasoline, LPG and CNG
PC connection
Via RS 232 port
Data-Logging
255 Tests
Dimensions
Weight
1kg
Handset
220mm x 55mm x 120mm
Probe
Insertion depth 35mm x Diameter 15mm
Clip handle to secure to exhaust, 4m long hose
Various probes available including high temperature
Ambient Operating Range
+5˚C to +45˚C/10% to 90% RH non condensing
Storage temperature
Minimum: 0˚C
Maximum: +50˚C
Battery Charger
Input: 110 Vac/230 Vac nominal
Output: 10 Vac off load
Analyzer battery run time
>4 hours form full charge with the pump running