Technical information

Technical information

Problem

Possible cause

Solution

No or low 

nebulisation 

rate

Nebulising button 

not pressed

Press the nebulising button on the 

nebuliser to start nebulisation. If the 

button is released, nebulisation will 

be interrupted.

Nebuliser held 

diagonally

Keep the nebuliser upright. Correct 

nebulisation can only be ensured if 

this positioning is used.

Atomiser not 

inserted or not 

correctly inserted

Correctly insert the atomiser (blue) 

into the medication container (see 

page 8).

Nozzle blocked

Switch off the device. Take the atom-

iser out of the medication container 

and check whether the nozzle below 

is blocked. Clean as necessary.

Air tube kinked or 

blocked

Check the throughput of the air tube. 

Clean or replace as required.

Air filter blocked

Clean air filter, replace as required.

Part no. 1802501, PZN-12399450, Medical aid code number: 14.99.99.1038

Original spare parts and accessories

•  AeroVital plus replacement kit, consisting of: 1 nebuliser, 1 air tube, 1 

adult mask, 1 child mask, 1 baby mask, 1 mouthpiece, 1 PEP training at-

tachment, 5 air filters

The following original spare parts and accessories are available from 

specialist dealers:

Customer service

The unit may only be repaired by the manufacturer or by an expressly au-

thorised organisation. Please contact:

The device does not require maintenance. Please note, however, the instruc-

tions for changing the air filter on page 12 and for cleaning and care of the 

device on page 19.

UEBE Medical GmbH 

Bgm.-Kuhn-Str. 22 

97900 Külsheim, Germany 

[email protected] 

www.uebe.com

Device of the protection class II

Please observe instructions for use

Keep away from sunlight

Storage and transportation conditions  

Ambient temperature -25 to +70 °C

Relative air humidity 10 to 95%

Keep dry

70 °C

-25 °C

95 %

10 %

Explanation of symbols

This product complies with the Council Directive 93/42/EC 

from 5 September 2007 regarding medical devices and bears 

the mark CE 0123 (TÜV SÜD Product Service GmbH).

Application part type BF: mouth piece and mask

0123

Содержание Domotherm vital plus

Страница 1: ...Instructions for use www domotherm de...

Страница 2: ...uliser where the medication is atomised into a fine mist that is easy to inhale Thank you for choosing the inhalation device Domotherm vital plus Domo therm vital plus is a compressor nebuliser and wa...

Страница 3: ...other accessories may cause malfunctions The warranty will be invalidated if the device is damaged by third party accessories Using this device in the vicinity of mobile phones microwaves or other dev...

Страница 4: ...outlet before cleaning Clean all necessary parts after each use as instructed in this guidebook See Page 19 Cleaning Cautions Unit description Air vault Angle adapter Air tube AeroVital plus nebulise...

Страница 5: ...ch the controls while seated Please note the instructions on cleaning and care from page 19 Fill the drug container with the amount intended for the treatment of the drug prescribed by your doctor Ima...

Страница 6: ...g on the side of the nebuliser unit will cause any drops that may have developed on the inside to run down so they can then be nebulised again The device has a PEP training function which can be used...

Страница 7: ...s with your doctor Should I use the mouthpiece or the mask for inhalation We recommend the mouthpiece for inhalation There is a more direct path to the lungs when breathing through the mouth than the...

Страница 8: ...f the nebulised me dium Here are a few examples of medication typically used for inhalation therapy Saline solution 0 9 Albuterole sulfate Ipratropium bromide Cromolyn sodium MMAD 1 3 m MMAD 2 3 m MMA...

Страница 9: ...arts and accessories AeroVital plus replacement kit consisting of 1 nebuliser 1 air tube 1 adult mask 1 child mask 1 baby mask 1 mouthpiece 1 PEP training at tachment 5 air filters The following origi...

Страница 10: ...ce and Accessories Wipe daily with a clean dry cloth only Medicine cup Air tubing Masks Mouth piece Device Disposable one time only X X X X X with water only Yes Yes Yes Yes Yes frequency of cleaning...

Страница 11: ...ble the nebulizer bottle when completely dry and put these parts in a dry sealed container Remove the air tube the nebuliser unit mouthpiece and mask Carefully unscrew the nebuliser unit Rinse the neb...

Страница 12: ...le power sources breakages and or dis assembly of the unit by the purchaser Furthermore no claims for damages against us are substantiated by the warranty Warranty claims can only be advanced in the w...

Страница 13: ...ternationally protected trademarks of Subject to technical modifications Not to be reproduced either in whole or in part Copyright 2022 UEBE Medical GmbH 7 18025 202 FA 2023 07 UEBE Medical GmbH Bgm K...

Отзывы: