Part 6: Final Sensor Assembly
Tools Needed:
Screwdriver, wire strippers
Components Needed:
spring-loaded assembly, terminal head assembly, terminal block, 2 screws
1. Insert the spring-loaded probe assembly into the head until the mounting plate rests against the connection head.
2. Feed the lead wires through the hole in the terminal block and slide the terminal block up the wires until it rests against
the mounting plate.
3. Screw the mounting screws loosely into the terminal head holding the terminal block and mounting plate.
4. Check the alignment of the assembly by pushing the housing into the head against the spring loading. When the
assembly is properly aligned, tighten the screws. Do not over tighten or you will crack the ceramic terminal block.
5. Cut and strip the red wire to the appropriate length and attach it to the negative terminal. Cut the other wire (or wires) to
the appropriate length and attach them to the positive terminal (terminals). (See Figure 5)
For thermocouple extension leads to controller, indicator or other device only use TC extension wire of the
same calibration.
Then, install the temperature sensor in your process!