The field to show measurements regarding the waveform  

2)

 

Indicates the horizontal trigger level  

3)

 

Indicates the vertical trigger level  

4)

 

Math function data 

5)

 

Exit from the oscilloscope mode 

6)

 

The level of vertical triggering, and which channel it stems from by its color 

7)

 

The ground level of Channel 1 

8)

 

The ground level of Channel 2 

9)

 

The ground level of Math Channel 

10)

 

Provides transition to the previous menu 

11)

 

The selection button for Channel 1 

12)

 

The setting button for Channel 1; vertical sensitivity, activation, coupling 

13)

 

The selection indicator for Channel 1 

14)

 

The selection button for Channel 2 

15)

 

The setting button for Channel 2; vertical sensitivity, activation, coupling 

16)

 

Horizontal sensitivity: Selecting sampling rate. (Time/div) 

17)

 

Trigger setting 

18)

 

Provides bmp formatted screen capture 

19)

 

Provides transition to the next menu 

20)

 

Automatically trigger or hold 

21)

 

Single triggering 

22)

 

Cursor is the menu where the observed signal is measured in terms of time and voltage 

23)

 

Enables oscilloscope to run as XY or XT option 

24)

 

Measure: Enables the requested measurement values that shown at  (1) to select as Time 

and/or Voltage 

25)

 

Math menu 

26)

 

Enables the data regarding to waveform to be recorded text formatted 

27)

 

Default settings 

 

3.2.

 

Trigger Setting 

To keep the observed signal constant on the display, the trigger source must be selected for that channel. 
Otherwise, it will be difficult to observe because the signal will be very moving on the display.  To choose 
the  trigger  source, trigger  menu should be  opened by pressing the  button shown as 17 in the  Figure  3a. 
Then, trigger channel and trigger edge should be selected from the menu seen in the Figure 4. 

Table 1. Trigger Modes 

CH1

 

Channel 1, rising edge 

CH1

 

Channel 1, falling edge 

CH2

 

Channel 2, rising edge 

CH2

 

Channel 2, falling edge 

EXT

 

External Trig, rising edge 

EXT

 

External Trig, falling edge 

Содержание UCE-DSO290

Страница 1: ...UCE DSO291 DIGITAL OSCILLOSCOPE USER MANUAL UCORE ELECTRONICS www ucore electronics com 2018...

Страница 2: ...Trigger Setting 6 3 3 Channel Setting 7 3 4 Horizontal Sensitivity Setting 8 3 5 Cursor Function 9 3 6 Auto Measurement 10 3 7 Operating Mode 11 3 8 XY Operating Mode 12 3 9 Mathematics Function 13 3...

Страница 3: ...mV div 10V div Horizontal sensitivity 10ns div 5s div Max input voltage 5V X1 Probe 50V X10 Probe Coupling AC DC Supply voltage 5V mini USB Trig modes Run single Hold functions Trigger functions risin...

Страница 4: ...turning off To turn on the device plug the USB cable into any USB socket whose output voltage is 5V Energy sources such as computer adapter and external power bank may be used as voltage source To tur...

Страница 5: ...cope Mode For accessing to oscilloscope mode DSO icon in the Main Menu should be selected The view of oscilloscope mode is seen in the Figure 2 Figure 2 Oscilloscope Mode overview 3 1 Display Descript...

Страница 6: ...www ucore electronics com 5 Figure 4b Oscilloscope mode Menu2 Figure 5c Oscilloscope mode Menu3...

Страница 7: ...s transition to the next menu 20 Automatically trigger or hold 21 Single triggering 22 Cursor is the menu where the observed signal is measured in terms of time and voltage 23 Enables oscilloscope to...

Страница 8: ...nto the point If the trigger will be made via external signal firstly the trigger should be selected from menu as an external channel Table 1 Henceforth signal generator output SG OUT is used as an ex...

Страница 9: ...ensitivity should be increased from the minimum level until the signal turns into observable state The vertical sensitivity may be different for the both channels however the horizontal sensitivity co...

Страница 10: ...buttons providing switching among the menus are used in the Figure 3a 10 and 19 buttons There are 4 cursors total which consists of 2 pcs t1 and t2 for measuring time 2 pcs V1 and V2 for voltage To ch...

Страница 11: ...minimum values it is calculated as the time of the remaining part over this level If it is out of the limit values then the average of the maximum and minimum level is taken and the duration of the r...

Страница 12: ...ime triggering is initiated necessarily If it forces occurrence of an invalid trigger the oscilloscope may not synchronize the wave form and waves are seen as rolling over in the display In case of a...

Страница 13: ...on the display gives an analog signature for the component The signature of all analog components is a combination of one or more of the four basic signatures resistance capacitance inductance and sem...

Страница 14: ...el 2 CH2 CH1 Subtraction function Channel 2 Channel 1 CH1 CH2 Multiplication function CH1 CH2 Division function Channel 1 Channel 2 CH2 CH1 Division function Channel 2 Channel 1 CH1 FFT Fast Fourier T...

Страница 15: ...reen capture it saves the current screen view as a 16 bit bmp formatted image via the Print Sc button numbered 18 2 Text formatted record of the data from ADC memory The values in ADC memory could be...

Страница 16: ...is between 0V 4V and output frequency is between from 1Hz to 50kHz with 1Hz resolution Signal generator s impedance is 50 Figure 13 Signal genarator On the left bottom of the display is a list of sig...

Страница 17: ...e 22 In order to change predefined colors on the left side of display click on it The new color is determined by clicking on the color scale on the right side Once the color personalization is complet...

Страница 18: ...s 7 File Browse The file browse menu is a menu where screen images recorded in the internal memory can be viewed again In the main menu by selecting Browse Files button users gain access to this menu...

Страница 19: ...fter four points if the calibration is successful the smiley face symbol appears on the display and the program returns to settings menu If the calibration is unsuccessful the program prompts you to g...

Страница 20: ...es connected to both channels are connected to GND terminal Figure 21 If the calibration parameters are stable within the specified limits touch anywhere on the display to complete the calibration and...

Отзывы: