8.2 Error Code

Error code display

Description

Err.

Calibration operation error

Err. 1

Counting setting error

Err. 2

The weighing pan is not placed correctly

Err. 3

Sample weight exceeds the capacity

Содержание USS-DBS83

Страница 1: ...Instruction Manual USS DBS83 Analytical Balance ...

Страница 2: ...ing Components 9 3 4 Leveling the Balance 9 3 5 Connecting the Power 11 Settings 12 4 1 Display Speed 12 4 2 Stability 13 4 3 Restore Factory Defaults 14 Calibration 15 Weighing and Applications 16 6 1 Weighing 16 6 2 Unit of Weight 17 6 3 Counting Function 18 6 4 Percent Weighing 20 Maintenance 21 7 1 Precautions 21 7 2 Cleaning 22 7 3 Disposal 22 Troubleshooting 23 8 1 Troubleshooting 23 8 2 Err...

Страница 3: ...se read the manual completely and follow the usage instructions before installation and operation as this will help you to make full use of the functions and performance of the USS DBS83 Analytical Balance 1 1 Safety Precautions The U S Solid USS DBS83 Analytical Balance qualifies as state of the art technology and complies with all recognized safety rules Improper use or handling however can resu...

Страница 4: ... type adapted to the specification of the balance If excessive voltage is used the balance may overheat and be damaged or cause a fire Operate the balance on a stable rigid and flat table Handle the balance carefully It is a precision device subjecting it to impact may result in a malfunction If the device is not be used for a long period of time the power should be turned off and disconnect the p...

Страница 5: ...3 Design and Function 2 1 Components 1 Weighing chamber 2 Weighing pan 3 Level bubble 4 Display panel 5 Keyboard 6 Glass door 7 Handle of glass door 8 Balance main body 9 Leveling foot ...

Страница 6: ...4 1 RS 232 connector 2 Power inlet ...

Страница 7: ... 2 2 Keyboard Key Description Turn the power on or off Enter settings and functions selection Tare Zero the balance Confirm selection Performs calibration procedure Switch between units Select mode item ...

Страница 8: ...ription 1 Weight value 2 Indicates positive values 3 Indicates negative values 4 Stabilization indicator 5 Capacity utilization ratio 6 Weight unit 7 Animal weighing 8 Automatic calibration time mode 9 Display speed 10 Stability ...

Страница 9: ...eteness The following accessories are part of the standard equipment for your new USS DBS83 Analytical Balance 1 Balance Main Body 1 Weighing Pan 1 Draft Shield 1 Power Cable 1 Calibration Weight 1 Instruction Manual 1 Quality Certification Check the instrument for damage in transit Immediately inform the U S Solid customer service if you have any complaints or parts are missing ...

Страница 10: ... 30 C the relative humidity is around 50 and that non condensing conditions are met Avoid direct sunlight and ensure that there are not any excessive temperature fluctuations Ensure the balance is places a sufficient distance from heat sensitive materials in the vicinity of the instrument Avoid the effects of air currents from air conditioners ventilators open doors or windows Keep away from objec...

Страница 11: ...r close 3 4 Leveling the Balance Accurate horizontal positioning necessary for repeatable measurements and exact results To compensate for small irregularities or tilts at this location the instrument needs to be leveled Adjust the leveling feet of the analytical balance until the air bubble in the indicator is centered The level indicator is located under the cover towards the rear of the analyti...

Страница 12: ...of air bubble Adjustment method up Turn both feet clockwise down Turn both feet counterclockwise Left Turn left foot counterclockwise right foot clockwise Right Turn left foot clockwise right foot counterclockwise ...

Страница 13: ...ource if it does not match Connecting a power cable a Insert the female end of the power cord into the power inlet located at the rear of the main unit b Plug the male end of the power cord into the outlet Note To obtain accurate results the analytical balance must be warmed up for at least one hour each time it is connected to an AC power source or after a power outage of more than 30 minutes Onl...

Страница 14: ...ons Description Highest speed High speed Normal speed Low speed Depending on the environment in which the instrument is used carefully select the display speed to avoid instability that might be caused by excessive speed 1 Press the SET button several times in succession until the SPEED is displayed 2 Press TARE to enter the selection of display speed mode 3 Press UNIT to select required speed and...

Страница 15: ...balance be adjusted appropriately in an environment that guarantees its normal operation Icons Description Lowest stability Low stability Normal stability High stability 1 Press the SET button several times in succession until the STAB is displayed 2 Press TARE to enter the selection of stability 3 Press UNIT to select required stability and confirm by pressing TARE ...

Страница 16: ...nstable the stability of the balance can also be adjusted by restoring the factory settings 1 Press the SET button several times in succession until the DEFAULT is displayed 2 Press TARE to confirm and an audible signal will sound 3 Then it will back to the splash screen the factory defaults have been restored The setting is only applicable to situations where settings are confusing due to incorre...

Страница 17: ... moving in the same room Changes in ambient conditions Prior to each use The balance must be fully warmed up for one hour to stabilize the weighing before calibration Procedure 1 Press TARE to zero the balance in the weighing state 2 Press the CAL button and the CAL 200 flashing displayed 3 Place the 200g calibration weight on the weighing pan 4 Then 200 0000g will be displayed External calibratio...

Страница 18: ...dicates a stable state 4 Read the value displayed Weighing a sample that needs to be in a container 1 Open the glass door of the weighing chamber and place the weighing container on the weighing pan then close the glass door again 2 Wait for the display to stabilize The appearance of the stability mark indicates a stable state Then press TARE to tare and the displayed value will return to zero 3 O...

Страница 19: ...an be used to meet the unit requirements in various usage situations 1 Press the UNIT button to select required unit 2 Press TARE to zero the displayed value and begin to weighing the sample To avoid the deviation of the weighing record caused by unit misreading please ensure that the selected unit is the one required ...

Страница 20: ...the counting function for small components it is necessary to ensure that the weight of all samples is consistent and that the weight of a single object is 0 5mg 1 Press the SET button several times in succession until the COU is displayed 2 Press TARE to enter the counting function COU 5PCS will be displayed 3 Press UNIT to select the quantity and put the corresponding quantity of samples Press T...

Страница 21: ... The operation can be started when the COU disappears and the corresponding number appears 5 After the test is finished press ON to select weight and press O T to confirm to exit and return to the weighing interface ...

Страница 22: ... the balance in the weighing state 2 Press the SET button several times in succession until PCT is displayed 3 Press TARE and PCT 100PCT will be displayed 4 Put the item to be used as a reference on the weighing pan and confirm by pressing TARE 5 100 00 PCT will be displayed Then remove the reference sample and put on the sample ...

Страница 23: ...height onto the weighing pan 3 Do not expose the balance to high humidity or dusty conditions for an extended period of time 4 When the balance is not to be used for a long time cover it with to prevent dust from entering 5 When weighing powder and fine particles should be done with a suitable container to prevent dust and particles from falling into the load cell below the weighing pan 6 Wear a g...

Страница 24: ...e balance using a piece of lint free cloth which has been wet with a mild detergent soap only Removing the sliding glass doors for cleaning 1 Remove the pan and anti draft ring from the weighing chamber 2 Unscrew and remove the inside knob on the glass door 3 Slide the glass door out backwards 7 3 Disposal Disposal of equipment and packaging must be carried out by the operator in accordance with r...

Страница 25: ...display is constantly changing or unstable Possible causes The sample pan is in contact with the draft shield or heating chamber Glass draft shield is not closed Test bench vibration Electromagnetic field interference Power supply instability Problem Displayed value and actual weight are not consistent Possible causes Not calibrated before weighing Not returned to zero before weighing Balance not ...

Страница 26: ...24 8 2 Error Code Error code display Description Err Calibration operation error Err 1 Counting setting error Err 2 The weighing pan is not placed correctly Err 3 Sample weight exceeds the capacity ...

Страница 27: ... 0 0001g 0 1mg Repeatability 0 0002g Linearity deviation 0 0003g Stabilization time Within 3 seconds Operating temperature 41 95 5 35 Pan size Φ80mm Interface RS232 Power 110V 60Hz AC Dimensions 11 8 7 9 11 6 inches 300 200 295 mm Net weight 13 23 lbs 6 kg Shipping weight 17 64 lbs 8 kg ...