Calibration
Calibration is a necessary step to assure the analytical balance will
accurately weigh the sample.
Perform calibration operations in the following situations:
•
Changes in the location of use (including moving in the same room).
•
Changes in ambient conditions.
•
Prior to each use.
The balance must be fully warmed up for one hour to stabilize the
weighing before calibration.
Procedure:
1. Press
“
TARE
”
to zero the balance in the weighing state.
2. Press the
“
CAL
”
button and the
“
CAL-200
”
flashing displayed.
3. Place the 200g calibration weight on the weighing pan.
4. Then
“
200.0000g
”
will be displayed. External calibration is completed.