Thank you for choosing U.S. Solid’s Analytical Balance. We hope you enjoy your pur-
chase. Please maintain the equipment according to all instructions within the manual.
This product was developed according to CE marking regulations while emphasizing
aesthetics and safety for the user. It has ISO certific tion, specifically ISO 9001 The
high quality of the product, coupled with proper user care, should allow you to enjoy
the product for years.
Improper use of this equipment can cause accidents, including but not limited to:
electric discharges, circuit breaker malfunctions, fi e, etc. Please read the mainte-
nance section, where this is explained more fully.
FOR BEST RESULTS AND LONGEVITY IN PRODUCT
EFFICACY, READ THROUGH THIS MANUAL THOROUGHLY
BEFORE USING THE PRODUCT
Please bear in mind the following:
•
This manual should be kept with the analytical balance, so it is available to all
users
•
Be careful when handling the balance. Avoid any sudden movements when in
its vicinity. Be sure to not drop the balance, nor drop any heavy/sharp objects
on it. Keep liquids away from the balance to avoid spills.
•
Never take apart the balance to try and fix ourself. This could ruin the entire
balance, and voids the product warranty. There is also risk of injury when
disassembling the balance.
•
To prevent fi e and electrical issues, avoid dusty or overly dry environments. In
case of electrical issue or fi e, unplug the balance immediately.
•
During setup, installation, and use, contact your wholesaler with any questions
that arise.
•
This equipment is protected under the Warranties and Consumer Good
Regulation (10/2003)
•
Overhaul is not covered by the equipment warranty
•
Changes made to the product by non-certified individuals will esult in a loss of
the product warranty
•
Accessories, including their loss, are not covered by the product warranty, nor
does the warrant cover piece’s deterioration over the course of time.
•
Keep your invoice, either for having the right to a warranty, or for asking for
warranty coverage. Manufacturer reserves the right to modify or improve the
manual or equipment.
I
I. Getting Started
Troubleshooting Guide
Problem
Cause
Solution
Display not working
No power supply
Plug in the adapter
Display shows
“
“
Object weight below
minimum threshold,
taring issues exist, or
balance is not properly
calibrated.
Make sure item being
weighed is at least 10mg.
Make sure receptacles
have been tared, and
the balance has been
properly calibrated.
Display shows
“
“
Item being weighed is
over maximum allow-
able weight.
Make sure balance is
properly calibrated, and
make sure item is below
maximum allowable
weight.
Display shows “Err-1”
Power issues with the
balance.
Machine may have a
poor power connection.
Check the power in the
back, and the outlet.
Leave machine off or
seveeral minutes, unplug
fully, then reconnect and
begin again.
Display shows “Err-2”
Unstable weighing con-
ditions.
Recalibrate machine,
make sure balance is
level, and make sure
item being weighed is
fully stable.