PAGE 9 OF 33
0218 IH-5848
MENU SETTINGS CONTINUED
SETUP MENU CONTINUED
AZT
– Set the automatic zero tracking functionality.
• OFF = Disabled.
• 0.5d = The display will maintain zero until a change
of 0.5 divisions per second has been exceeded.
• 1d = The display will maintain zero until a change of
1 divisions per second has been exceeded.
• 3d = The display will maintain zero until a change of
3 divisions per second has been exceeded.
StAbLE
– Set the amount the reading can vary while the
stability symbol remains on.
• 0.5d = 0.5 balance division.
• 1d = 1 balance division.
• 2d = 2 balance division.
• 5d = 5 balance division.
L IGHT
– Sets backlight functionality.
• OFF = Always off.
• ON = Always on.
• AUTO = Turn on when a button is pressed or the
displayed weight changes.
NOTE: When connected with power pack, the
backlight is always on.
A.tArE
– Set the automatic tare functionality.
• OFF = Automatic Tare is disabled.
• ON = The first stable gross weight is tared.
• ON-ACC = Stable gross loads within the accept
limits are tared (in Check weighing mode).
A.OFF
– Set the automatic shut off functionality.
• OFF = Disabled
• 1 = Powers off after 1 minute of no activity.
• 5 = Powers off after 5 minutes of no activity.
• 10 = Powers off after 10 minutes of no activity.
End
– Advance to the next menu or return to the top of
the current menu.
MODE MENU
This menu activates modes so they will be available to
use with the
MODE
button.
RESET
no,
yes
WEIGH
off,
on
COUNT
off,
on
PERCENT
off,
on
CHECK
off,
on
TOTALIZE
off,
on
HOLD
Disp,
Peak, Off
END MODE
exit menu
rESET
– Reset the Mode menu to factory defaults.
• NO = Not reset.
• YES = Reset.
WEIGH
– Set the status.
• OFF = Disabled.
• ON = Enabled.
COUNt
– Set the status.
• OFF = Disabled.
• ON = Enabled.
PErCNt
– Set the status.
• OFF = Disabled.
• ON = Enabled.
CHECK
– Set the sub-mode.
• OFF = Disabled.
• ON = Enabled.
totAL
– Set the sub-mode.
• OFF = Disabled.
• ON = Enabled.
HoLd
– Set the sub-mode.
• OFF = Disabled.
PEAK HOLD = allows the user to capture and store the
highest stable weight value (>=5d)
DISPLAY HOLD = allows the user to capture and store the
first stable weight value (>=5d)
When the stable value is being held on the display, the
"Hold" icon will blink and the displayed weight will not
change.
End
– Advance to the next menu or return to the top of