PAGE 4 OF 32
0521 IH-2289
EXPONENT
Set the factor multiplier:
0
= 10
0
(Factor x 1)
1
= 10
1
(Factor x 10)
2
= 10
2
(Factor x 100)
3
= 10
3
(Factor x 1,000)
-2
= 10
-2
(Factor 100)
-1
= 10
-1
(Factor 10)
Least Significant Digit
:
Set the custom unit readability:
0.00001, 0.00002, 0.00005, 0.0001,
0.0002, 0.0005, 0.001, 0.002, 0.005,
0.01, 0.02, 0.05, 0.1, 0.2, 0.5, 1, 2, 5,
10, 20, 50, 100, 200, 500, 1,000
NOTE: LSD settings are limited to
values that result in a displayed
resolution of 1,000 to 30,000
divisions.
END UNIT
Advance to the next menu.
CALIBRATION MENU
Three calibration processes are available:
zero calibration, span calibration and linearity
calibration. (See Figure 2)
1. Make sure that appropriate calibration masses are
available before beginning calibration.
2. Make sure that the scale base is level and stable
during the entire calibration process.
3. Calibration is unavailable with LFT set to ON.
4. Allow the Indicator to warm up for approximately five
minutes after stabilizing to room temperature.
5. To abort calibration, press the
EXIT
button anytime
during the calibration process.
6. When any selection within the GMP menu is enabled,
calibration results are automatically printed.
ZERO CALIBRATION
Zero calibration uses one calibration point. The zero
calibration point is established with no weight on
the scale. Use this calibration method to adjust for a
different pre-load without affecting the span or linearity
calibration.
1. When
ZErO
is displayed, press the
YES
button to initiate zero calibration.
2. The display flashes
0
and the
calibration unit. Press the
YES
button
to establish the zero point.
3. The display shows
--C--
while the
zero point is established.
4. When zero calibration is completed,
the display shows
dONE
.
5. The scale exits to the active
weighing mode and displays the
actual weight value.
SPAN CALIBRATION
Span calibration uses two points to adjust the scale. The
span calibration point is established with a calibration
mass placed on the scale. The zero calibration point is
established with no weight on the scale.
1. When
SPAN
is displayed, press
the
YES
button to initiate span
calibration.
2. The display flashes the span
calibration point. Place the
specified weight on the scale and
press the
YES
button.
3. To choose a different span point
or calibration unit, edit the setting
as explained in Menu Navigation.
When the desired setting is displayed,
place the specified weight on the
scale and press the
YES
button.
4. The display shows
--C--
while the
span point is established.
OPERATION CONTINUED
Figure 2
Zero Perform
Span Perform
Linearity Perform
Cal Test
Perform
Geographic
Adjustment
Set 00...
Set 12
... Set 31
End Calibration Exit CALIBRATE menu
ZErO
--C--
0
kg
CALIBRATION VALUES
MODEL #
SPAN CALIBRATION
POINTS (kg.)
LINEARITY CALIBRATION
POINTS (kg.)
H-2289
50
25, 50
H-2290
100
50, 100
H-2291
250
125, 250
•
•
•
•
•
•
LSd
C
End.UN
dONE
. O.000
kg
SPAN
30
kg
25
kg
--C--