0422 IH-9884

TROUBLESHOOTING

TECHNICAL DATA

OPERATING ISSUE

POSSIBLE CAUSE

REMEDY

Cannot turn on

No power to scale.

Verify connections and voltage.

Poor accuracy

Improper calibration. 

Unstable environment.

Perform calibration. 

Move scale to suitable location.

Cannot calibrate

Unstable environment. 

Incorrect calibration weight.

Move scale to suitable location. 

Use correct calibration weight.

Cannot access mode

Mode not enabled.

Enter menu and enable mode.

Cannot access unit

Unit not enabled.

Enter menu and enable unit.

Lo rEF

Reference weight is too low.

Increase reference weight.

rEF Err

Parts counting – sample weight < 1d.

Shows error, exits mode or goes to 

Clr.APU.

Err 3.0 CAL

Incorrect calibration weight.

See Calibration on page 3.

ERR 4.4 FuLL

RS232 Buffer is full. 

Set Handshake on, see Interface User Manual.

Err 8.1 LoAd

Power on zero range exceeded.

Clear pan; check shipping lock setting.

Err 8.2 LoAd

Power on zero under range.

Install pan; check shipping lock setting.

Err 8.3 LoAd

Overload.

Load exceeds scale maximum capacity.

Err 8.4 LoAd

Under load.

Reading below min. range; re-install pan.

Err 8.6 999999

Displayed value > 999999.

Result exceeds display capability.

Err 9 dAtA

Internal data error.

Contact Uline Customer Service.

Err 13

 

MEM

Fail to write EEPROM.

Contact Uline Customer Service.

Err 53 CSUM

Invalid checksum data.

Contact Uline Customer Service.

If troubleshooting section does not resolve or describe problem, contact Uline Customer Service at 1-800-295-5510.

EQUIPMENT RATINGS

•  Pollution degree 2

•  Installation category II

•  Altitude 2000m

•  Humidity: Maximum 80% for temperatures up 

to 88˚F decreasing linearly to 50% at 104˚F 

non-condensing

•  Electrical supply: Rated 12VDC 500mA for use with a 

certified/listed power adapter, or battery operated

•  Indoor use only

•  Temperature range: 50˚F to 104˚F

•  The main supply voltage fluctuations are not to 

exceed ± 10% of the nominal supply voltage.

Содержание H-9884

Страница 1: ...e on Long Press when on Turns the scale off Short Press in Menu Selects accepts displayed setting Long Press Toggles through active units Short Press in Menu Toggles through available settings Short P...

Страница 2: ...Disconnect from power supply when cleaning Service should only be performed by authorized personnel SETUP TRANSPORTATION LOCK 1 Rotate the pointer on the transportation lock to the unlocked position S...

Страница 3: ...4 After pressing yes the scale will then display the stored sample size Press NO or BACK to toggle the choices 5 10 20 50 or 100 5 Put the indicated number of pieces on the pan Press YES to calculate...

Страница 4: ...on procedure zero and span A span calibration is important when initially setting up the scale Lin LIN yes no Initiates a linearity calibration procedure zero mid point and span SETUP SUB MENU Back Li...

Страница 5: ...he menu must be turned on to be active NOTE Available units and modes vary by model and local regulations SECURITY SWITCH Menu Lock switch may be used to limit changes to the user menu See Figure 4 Th...

Страница 6: ...nge exceeded Clear pan check shipping lock setting Err 8 2 LoAd Power on zero under range Install pan check shipping lock setting Err 8 3 LoAd Overload Load exceeds scale maximum capacity Err 8 4 LoAd...

Страница 7: ...des Weighing Parts Counting Checkweighing Power Requirement AC adapter included or four C batteries not included Typical Battery Life 270 hours 200 hours 270 hours Specified Temperature Range 50 F to...

Страница 8: ...0 0001 21 574 x 0 001 Ounce oz 7 7600 x 0 0005 77 6025 x 0 0005 77 600 x 0 005 Ounce Troy ozt 7 0730 x 0 0005 70 7315 x 0 0005 70 730 x 0 005 Pennyweight dwt 141 46 x 0 01 1 414 63 x 0 01 1 414 6 x 0...

Отзывы: