PAGE 6 OF 8
0422 IH-9884
TROUBLESHOOTING
TECHNICAL DATA
OPERATING ISSUE
POSSIBLE CAUSE
REMEDY
Cannot turn on
No power to scale.
Verify connections and voltage.
Poor accuracy
Improper calibration.
Unstable environment.
Perform calibration.
Move scale to suitable location.
Cannot calibrate
Unstable environment.
Incorrect calibration weight.
Move scale to suitable location.
Use correct calibration weight.
Cannot access mode
Mode not enabled.
Enter menu and enable mode.
Cannot access unit
Unit not enabled.
Enter menu and enable unit.
Lo rEF
Reference weight is too low.
Increase reference weight.
rEF Err
Parts counting – sample weight < 1d.
Shows error, exits mode or goes to
Clr.APU.
Err 3.0 CAL
Incorrect calibration weight.
See Calibration on page 3.
ERR 4.4 FuLL
RS232 Buffer is full.
Set Handshake on, see Interface User Manual.
Err 8.1 LoAd
Power on zero range exceeded.
Clear pan; check shipping lock setting.
Err 8.2 LoAd
Power on zero under range.
Install pan; check shipping lock setting.
Err 8.3 LoAd
Overload.
Load exceeds scale maximum capacity.
Err 8.4 LoAd
Under load.
Reading below min. range; re-install pan.
Err 8.6 999999
Displayed value > 999999.
Result exceeds display capability.
Err 9 dAtA
Internal data error.
Contact Uline Customer Service.
Err 13
MEM
Fail to write EEPROM.
Contact Uline Customer Service.
Err 53 CSUM
Invalid checksum data.
Contact Uline Customer Service.
If troubleshooting section does not resolve or describe problem, contact Uline Customer Service at 1-800-295-5510.
EQUIPMENT RATINGS
• Pollution degree 2
• Installation category II
• Altitude 2000m
• Humidity: Maximum 80% for temperatures up
to 88˚F decreasing linearly to 50% at 104˚F
non-condensing
• Electrical supply: Rated 12VDC 500mA for use with a
certified/listed power adapter, or battery operated
• Indoor use only
• Temperature range: 50˚F to 104˚F
• The main supply voltage fluctuations are not to
exceed ± 10% of the nominal supply voltage.