OPERATION CONTINUED
Head
er0
Head
er1
Head
er2
Head
er3
Weig
ht1
Weig
ht2
Weig
ht3
Weig
ht4
Weig
ht5
Weig
ht6 Tare1 Tare2 Tare3 Tare4 Tare5 Tare6
Termin
ator1
Termin
ator2
Con2:
Header0=02H
Header1 follow decimal point
Decimal point=0, header1=22H
Decimal point=1, header1=23H
Decimal point=2, header1=24H
Decimal point=3, header1=25H
Decimal point=4, header1=26H
Header2 follow weigh status, default value=20H
If in net mode (tare value not 0), header2=header2|01H
If gross weight "-", header2=header2|02H
If overload or gross weight "-", header2=header2|04H
If unstable, header2=header2|08H
If weighing unit=kg, header2=header2|10H
Header3 follow weighing unit
If weighing unit=g, header3=21H
If weighing unit=oz, header3=23H
Weight1~weight6: weighing data
Tare1~tare6: tare value
Terminator1: 0DH
Terminator2: 0AH
Con3
:
Header
0
Header
1
Weight
1
Weight
2
Weight
3
Weight
4
Weight
5
Weight
6
Weight
7
Unit1
Unit2 Status Termin
ator1
Termin
ator2
HEADER0=01H
Header1 follow weight "+" or "-"
When weight "+", header1="+", when weight "-", header="-"
Weight1~weight7: weight data (include decimal point)
Unit1~unit2: weight unit
Status: when stable, status=0, when unstable, status=1
Terminator1: 0DH
Terminator2: 0AH
LEGAL FOR TRADE
When the indicator is used in trade or a legally
controlled application, it must be set up, verified and
sealed in accordance with local weights and measures
regulations. It is the responsibility of the purchaser to
ensure that all pertinent legal requirements are met.
CALIBRATION
1. Break the seal on the indicator housing.
2. Open the housing.
3. See calibration process on page 5. Complete steps 1-5.
4. After step 5, press the calibration switch on the
mainboard. (See Figure 10)
5. Complete remaining calibration steps on page 5.
VERIFICATION
The local weights and measures official or authorized
service agent must perform the verification procedure.
SEALING
The local weights and measures official or authorized
service agent must apply a security seal to prevent
tampering with the settings.
Top
Calibration Switch
Figure 10
SETUP MENU
BACKLIGHT
At startup, the backlight will automatically turn on
when the scale is in use. Enter this menu to change the
backlight setting.
1. Press and hold the
ZERO
key
for 3 seconds.
SET bF
will be shown on the display.
2. Press the
ZERO
key
to enter the backlight menu.
3. Press the
TARE
key
until the desired setting is
shown on the display.
•
AU
– Backlight automatically turns on when the
scale is in use.
•
on
– Backlight is always on.
•
oFF
– Backlight is never on.
4. Press the
ZERO
key
to select the new setting.
5. Press the
UNIT
key
to exit the sub menu.