0521 IH-2587
If the troubleshooting section does not resolve your problem, contact Uline Customer Service at 1-800-295-5510.
NOTE: Before using the indicator for the first time, the internal rechargeable battery should be fully
charged for 12 hours. The indicator can be operated during the charging process. The battery is
protected against overcharging and the indicator can remain connected to the AC power line.
TROUBLESHOOTING CONTINUED
OPERATING ISSUE
CAUSES
RECOMMENDATIONS
Error 8.1
Weight reading exceeds Power On Zero
limit.
Remove load from scale. Recalibrate
scale.
Error 8.2
Weight reading below Power On Zero limit.
Remove red shipping tags from under pan.
Add load to scale. Recalibrate scale.
Error 8.3
Weight reading exceeds Overload limit.
Reduce load on scale.
Error 8.4
Weight reading below Underload limit.
Add load to scale. Recalibrate scale.
Err 9.0
Internal fault.
Service required.
Err 9.5
Calibration data not present.
Calibrate scale.
Err 53
EEPROM data incorrect.
Service required.
CAL E
Calibration Error. Calibration value outside
allowable limits.
Repeat calibration using correct
calibration weights.
LOW.rEF
The average piece weight of the parts is
small (warning).
Use parts with average piece weight
greater than or equal to 1 division.
REF.WT Err
The average piece weight of the parts is
too small.
Use parts with average piece weight
greater than or equal to 0.1 division.