3
. A beep will sound and the code number will appear fol-
lowed by a flashing “ ” symbol. Be sure the code number
on the display matches the code number on the Test Strip
vial. If the code numbers do not match, code the Meter
by using Glucode Chip, accordingly.
4
. After the “ ” symbol appears, press the
S
button for 3 sec
to switch “ctl” mode. The “ctl” symbol indicates that the
Meter will mark your current test as a Control Solution
test. Now you are ready to apply the Control Solution.
5
. Discard the first drop of Control Solution and squeeze a
small drop of Control Solution on a clean nonabsorbent
surface (such as a clean piece of wax paper).
Do not
apply Control Solution to the Test Strip directly from the
bottle.
6
. Bring the tip of the Test Strip to lightly touch the drop of
Control Solution. Control Solution is automatically pulled
into the strip through the tip. Hold until the meter beeps.
The meter will now being counting down from 8 to 1 and
the Control Solution test result will appear.
AM
mg/dL
ctl
18
M
S