8.10
Opening the door and Unloading
1. Push the
button to confirm the "Successful" or the
"error" message to unlock the door.
2. Open the door.
Caution!
Open the door a little to release the steam from the chamber. Only
after the steam escaped, open the door widely.
Warning!
To avoid severe injuries from hot steam and condensed hot water
that may drip out when opening the door, it is strictly forbidden to
lean on the autoclave. It is strictly forbidden to place your hand or
any part of your body over or under the door.
3. Use the tray handle or wear heat-resistant gloves to remove the
load from the autoclave.
4. After unloading, visually inspect the load to ascertain that it is dry,
and that sterilization indicators have made the required color
change.
Caution:
Water droplets and visible signs of moisture on sterile packaging
or the tape used to secure it, may compromise sterility of
processed loads or be indicative of a sterilization process failure.
Visually check outside wrapper for dryness. If there are water
droplets or visible moisture on the exterior of the package or on
the tape used to secure it, the pack or instrument tray is
considered unacceptable.
Page 60