Addendum: Clean-In-Place System
E09 TD-4100XD
A-10
E09 TD-4100XD Oil In Water Monitor
Clean-In-Place Addendum
PN 106169
Rev B
Select “1” to set the starting date (must be a current or
future date).
Select “2” to set the starting time (must be a future time).
You must also set the cleaning frequency. Return to the
Cell Cleaning screen, and press “4.” You must set the
interval of the cleaning as well as the units. Use the “→”
key to change the units.
Finally, you must set the cleaning sequence. This is done
by pressing “5” at the Cell Cleaning screen.
Press “1” to set the duration for keeping the inlet valve
closed. Use the keypad to enter a value in seconds.
All settings for the cleaning cycle are entered in seconds.
Press “2” through “5” to enter duration values for the
corresponding steps in the cleaning cycle.
Always allow more time for draining the system than
cycling the cleaning solution.