SPECIFICATION
Notes:
1) Based on ITS-90 ( ITS: The International Temperature Scale); the environmental
condition: (23±5)℃, ≤85%RH; it should be put in the stable environmental conditions for at
least one hour, wait 5 minutes after start-up, with the menu parameters of S_rAtE=1 and
FILt=1, exclusive of sensor errors. The tolerance is 1.2Δ in case of S_rAtE=2.
2) Using the method of “Calibration” in page 12 of this manual, the fixed error component
between sensor and thermometer can be eliminated.
3) The thermal resistance and thermocouple’s measurement range and tolerance
shown in other units (Ω, mV,
℉
and K) are equivalent to the above table.
PAGE -06