Calibration adjustment of TRPG -01B is performed electronically via internal software
with case closed. There are no mechanical adjustments. All commands and adjustments
are done via the key pad using display to guide the user through calibration
procedure
CALIBRATION
PRESSURE CALIBRATION
(DEAD WEIGHT PRESSURE SOURCE RECOMMENDED)
Switch on the gauge & immediately Press the
Password ‘00000’ (factory set password) is
Press enter key, the gauge will enter
CALI’ is displayed.
There are total 8 steps for Pressure Calibration.
Fn key.
displayed.
calibration mode
Press Enter
The pressure is to be applied in “kg/cm ” only.
Apply the displayed pressure and press enter.
Observe the sensor output & press enter after it is steady.
Repeat the above steps for all the displayed eight pressure points.
The procedure will come to an end after pressing enter on the last pressure point.
Press Fn key to return to the gauge to normal operation & validate the calibration.