26
Cleaning Your Lancing Device
Cleaning Procedures
1. Take out one alcohol swab with 70% alcohol from the package and
squeeze out any excess liquid in order to prevent damage to the
meter.
2. Wipe all lancing device's exterior surface and release button.
3. Remove the alcohol swab. Allow the lancing device surface to dry
completely.
4. Discard the used alcohol swabs and never reuse them. Users
should wash hands thoroughly with soap and water after handling
the lancing device.
The lancing device should be cleaned after each use.
Important Control Solution Information
§
Use only Tula Health control solutions with your meter.
§
Do not use the control solution beyond the expiration date or 3
months after first opening. Write the opening date on the control
solution vial and discard the remaining solution after 3 months.
§
It is recommended that the control solution test be done at room
temperature 68°F to 77°F (20°C to 25°C). Make sure your control
solution, meter, and test strips are at this specified temperature
range before testing.
§
Shake the vial before use, discard the first drop of control solution,
and wipe off the dispenser tip to ensure a pure sample and an
accurate result.
§
Store the control solution tightly closed at temperatures between
35.6°F to 86°F (2°C to 30°C). Do NOT freeze.