Reviewing Your Test Results
The GlucoPet meter saves up to 200 test results in its built-in memory
along with the average glucose level for the number of Days that you set.
This enables you to recall the history of tests or average results for a
particular period at the time. The newest result will replace the oldest
if there are more than 200 tests in the memory. Enter “Reviewing Your
Test Results” mode by pressing the MEM button.
To delete the test results individually, press the SET button for 5 seconds
until you hear a “beep” sound. The LCD window will show the test result
and time deleted.
Delete Results
Display Past Results
To display the “Past Results” mode, press the SET button to review
saved results, beginning with the most recent set date and time.
IMPORTANT
: If you do not set the meter properly, the average glucose
level for the number of days you set and the results in the memory
will not show the proper value, but will indicate an error message.
An incorrect date setting will produce an “Er3” message on the LCD
window.
Blood applies to the top edge of the test strip is automatically drawn into
the channel where the reaction takes place.
NOTE:
Do not drop blood or control solution on the strip.
GlucoPet Test Strips
Check here to confirm that
sufficient blood has been applied.
Insert this end into the meter.
Push it in until it will go no
further.
Apply blood here
Step 1:
Insert test strip in meter. The meter will turn on.
How to Perform a Blood Glucose Test
Step
2:
Match the code number on the Meter LCD window
with the code number on the test strip vial by pressing the
set Button.
Note
: If the LCD window shows no code or the code number
disappears before you see the code number on the vial, remove
the test strip and try again by firmly inserting the test strip into
the test port.
Step 3:
Prick recommended place for your animal.
See Website: www.animaldiabetes.com
Step 4:
Pick up meter with test strip in it, and gently touch
top of strip at a 45 degree angle down into the blood drop
until you hear the“Beep” sound.
Step 5:
Your result will appear in 5 seconds.
Simple to Use
Step 1:
Insert test strip in meter. The meter will turn on.
.
ERROR 2
Using polluted test strip and used strip.
WHAT TO DO
With a new test strip, re-test.
ERROR 3
Poor amount of blood sample.
WHAT TO DO
With a new test strip and sufficient blood sample,
re-test.
Er 4
Monday mg/dl
6 12 12:30PM
Er 3
Monday mg/dl
6 12 12:30PM
Er 2
Monday mg/dl
6 12 12:30PM
Er 1
Monday mg/dl
6 12 12:30PM
Er 5
Monday mg/dl
6 12 12:30PM
ERROR 4
Problem with the test strip.
WHAT TO DO
With a new test strip, re-test.
ERROR 5
This message appears when blood sample is input
before the symbol for blood blinks.
WHAT TO DO
Wait for the symbol to blink then re-test using
Proper amount of blood.
ERROR
Problem with the meter
WHAT TO DO
If the problem persists, contact ADM at 866-391-9436.
Error Messages
HI
Monday mg/dl
6 12 12:30PM
Lo
Monday mg/dl
6 12 12:30PM
HI Reading
The test result is higher than 600mg/dl.
WHAT TO DO
Re-test with blood samples 2 or 3 times. “Hi”
result indicates severe hyperglycemia(High blood
glucose). If the “Hi” message persists, seek medical
attention immediately.
Lo Reading
The test result is lower than 10mg/dl.
WHAT TO DO
Re-test with blood samples 2 or 3 times. “Lo”
result indicates severe hypoglycemia (Low
blood glucose). If the “Lo” message persists, seek
medical attention immediately.
Changing the Batteries
Your meter comes with one 3.0V (CR3032 or equivalent)
lithium battery. The battery will provide you with sufficient
power to perform about 5000 tests. When replacing the
battery, only a 3.0V lithium battery may be used.
Lifetime Warranty
If the meter malfunctions for any reason, with the
exception of obvious abuse, misuse or damage, ADM will
replace it free of charge.
Please complete the Warranty Registration Card supplied
with the meter and fax or mail it to ADM. You can also
register your meter online at www.animaldiabetes.com.
ERROR MESSAGES
HI C Reading
The ambient temperature is to high.
WHAT TO DO
Place the meter in the operating range of
temperature for more than 10 minutes then re-test.
Lo C Reading
The ambient temperature is too low.
WHAT TO DO
Place the meter in the operating range of
temperature for more than 10 minutes then re-test.
Lo C
Monday mg/dl
6 12 12:30PM
Monday mg/dl
6 12 12:30PM
HI C