Tuffcare Rhino P440 Скачать руководство пользователя страница 4

6

7

RHINO STAND UP PATIENT LIFTER P440 and P445

RHINO STAND UP PATIENT LIFTER P440 and P445

WEIGHT LIMIT

DO NOT

 exceed maximum weight limit of the patient lifter. 

The weight limit for the stand up lifter is 350 lbs.

ASSEMBLING THE LIFTER

DO NOT

  over-tightens mounting hardware. It will cause a 

damage of mounting brackets.

USING THE SLING

Be sure that sling is properly attached on sling holders before 
the patient is removed from a bed, chair or any objects.
If the patient is in a wheelchair, secure the wheel locks in place 
to prevent the chair from moving forwards or backwards.

OPERATING THE LIFTER

Prior to first, make sure that battery is being fully charged. 
Always use the handles to move the lifter.

LIFTING THE PATIENT

Before positioning the legs of the stand-up lifter around the 
patient, make sure that the patient’s feet are out of the way of 
the foot plate, otherwise injury may occur.

Adjustments for safety and comfort should be made before 
moving the patient. Patient’s arms should be outside of the 
sling straps.

Before lifting a patient from a wheelchair, bed or any objects, 
slightly raise the patient off the object and check that sling 
attachments are secured. 

During transfer, with the patient suspended in a sling attached 
to the lifter, 

DO NOT

 roll caster base over uneven surfaces that

would create an imbalance of the patient lifter and could 
cause the patient lifter to tip over. Uses handle on the mast at 
all times to move the patient lifter.

GENERAL GUIDELINES 

contains important information for the 

safe operation and use of this Medical Device.

SECTION 1

GENERAL GUIDELINES

Check all parts for shipping damage before using. In 
case of damage,

 DO NOT

 use the equipment. 

The 

TUFFCARE

 patient lifter is NOT a transport 

device. It is intended to transfer a patient from 
one seated surface to another (such as a bed to a 
wheelchair).

DO NOT

 attempt to transfer, the user must be 

assessed by a qualified professional. Thoroughly 
read and fully understand the instructions in this 
Owner’s Manual,

TUFFCARE

 Stand up and Transfer slings are 

specifically designed to be used in conjunction with 

TUFFCARE

 patient lifter. Slings and accessories 

designed by other manufacturers are not to be 
utilized as a component of Tuffcare’s patient lifter 
system. 

The stand up lifter is intended to be used for patient 
within weight limit indicated for the lifter. 

DO 

NOT

 attempt to lifter more than the weight limit 

indicated.

1

2

3

4

5

WARNING

Содержание Rhino P440

Страница 1: ......

Страница 2: ...he Patient Transferring to a Wheelchair Transferring to a Bed Maintenance Safety Inspection Checklist Cleaning the Lifter Detecting Wear and Damage Lubricating the Lifter Warranty SECTION 3 Operation...

Страница 3: ...ctor and a grounding plug The plug MUST be plugged into an outlet that Is properly Installed and grounded in accordance with all local codes and ordinances If you ever feel even a slight shock from th...

Страница 4: ...the object and check that sling attachments are secured During transfer with the patient suspended in a sling attached to the lifter DO NOT roll caster base over uneven surfaces that would create an i...

Страница 5: ...aintenance of patient lifter and accessories is required to assure proper operation Casters and axle bolts are also required to be checked for tightness DO NOT over tightens mounting hardware It will...

Страница 6: ...les on the bracket on the mast Check and tighten all hardware BEFORE use Use both hands pull both adjustment pins outward at the same time Position the knee pad to the desired height on the mast and r...

Страница 7: ...the battery case on control unit at the desired position Push the top of battery case against mounting bracket until there is an audible click PUSH HANDLE Place the control unit on the mounting bracke...

Страница 8: ...e connections on the bottom of the control unit Check and tighten all hardware BEFORE use Control Unit Plug for Power Cord Plug for Hi low Actuator Plug for Pendant Plug for Leg Actuator For P445 only...

Страница 9: ...tton clockwise EMERGENCY RELEASE To activate the emergency release push the Red Button powered by 9V battery located on the side of Hi low actuator to lowering down the lift arm Open the battery cover...

Страница 10: ...is off and the unit will not function or charge 3 Plug the power code into a 110 Volt outlet NOTE When plugged into an AC outlet the safe protect feature turns on disabling the control pendent from f...

Страница 11: ...s on the lift arm Check that the patient sling is properly connected to the attachment holders BEFORE lifting the patient During transfer with the patient suspended in the patient sling DO NOT roll th...

Страница 12: ...ructions in this Owner s Manual ALWAYS keep hands and fingers clear of moving parts to avoid injury Individuals that use the stand up patient sling MUST be able to support their own weight otherwise i...

Страница 13: ...bottom edge of the transfer sling is at the base of the patient s spine 4 The rear casters are unlocked 5 The legs are in the maximum open position Instruct the patient to hold onto the lift arms on b...

Страница 14: ...r optimum stability and safety Be sure to check the sling attachments each time the sling is removed and replaced to ensure that it is properly attached before the patient is removed from a surface TR...

Страница 15: ...ise the patient above the surface of the bed The patient should be elevated just high enough to clear the bed with their weight fully supported by the lifter 2 Press the DOWN button and lower the pati...

Страница 16: ...at Tuffcare option Tuffcare sole obligation and your exclusive remedy under this warranty shall be limited to such repair and or replacement This warranty does not include any labor charges incurred i...

Страница 17: ......

Отзывы: