Changing the Power Supply Fuses
CAUTION
! Only fuses with the required rated current, voltage, and specified type should
be used with this device. Use only 5 x 20mm, 250V 1.6A rated, slow-blow fuses.
To change the power supply fuses:
1.
Turn off the power switch. Disconnect the power cord from the power supply.
2.
Using a small flathead screwdriver gently push the clip along the left side of the fuse
plate to the right.
3.
When the fuse plate pops free, gently slide the fuse plate and housing out of the power
supply.
4.
The housing contains two fuses. Replace the defective/broken fuse(s) with a new 5 x
20mm, 250V 1.6A rated, slow-blow fuse by gently pushing the end of the fuse into the
fuse housing.
5.
Push the fuse housing back into the power supply and press gently until the clip clicks
into place.
Cleaning the RZ2 Processor Device
To clean the device:
1.
Remove power from the RZ2 device chassis.
2.
Clean the external surfaces and LCD screen of the device with a soft, dry cloth.
3.
Do not attempt to disassemble and clean the inside of the device.