Changing the Power Supply Fuses  

CAUTION

! Only fuses with the required rated current, voltage, and specified type should 

be used with this device. Use only 5 x 20mm, 250V 1.6A rated, slow-blow fuses.  

 

To change the power supply fuses:  

1.

 

Turn off the power switch. Disconnect the power cord from the power supply.  

2.

 

Using a small flathead screwdriver gently push the clip along the left side of the fuse 
plate to the right.  
 

                  

 

3.

 

When the fuse plate pops free, gently slide the fuse plate and housing out of the power 
supply.  

4.

 

The housing contains two fuses. Replace the defective/broken fuse(s) with a new 5 x 
20mm, 250V 1.6A rated, slow-blow fuse by gently pushing the end of the fuse into the 
fuse housing.  

5.

 

Push the fuse housing back into the power supply and press gently until the clip clicks 
into place. 

 

Cleaning the RZ2 Processor Device 

To clean the device: 

1.

 

Remove power from the RZ2 device chassis. 

2.

 

Clean the external surfaces and LCD screen of the device with a soft, dry cloth. 

3.

 

Do not attempt to disassemble and clean the inside of the device. 

Push Fuse 
Clip to the 
right. 

Содержание RZ2

Страница 1: ...RZ2 BioAmp Processor Operator s Manual ...

Страница 2: ...s is and is subject to being changed without notice TDT shall not be liable for errors or damages in connection with the furnishing use or performance of this document or of any information contained herein The latest versions of TDT documents are always online at www tdt com support htm A CAUTION informs users when failure to take or avoid a specified action could result in damage to the product ...

Страница 3: ...building on the strengths of its predecessors while improving flexibility and overall processing power Functional Diagram The RZ2 device chassis houses an integrated zBUS communication interface and power source Power Supply The RZ2 s power supply is integrated into the device and is shipped from the factory with the desired voltage setting 115 V or 230V preset If you need to change the voltage se...

Страница 4: ...3 Turn the power switch on and check that the power switch s blue LED is illuminated Disconnecting Power from the RZ2 CAUTION When removing the power cord from either the power supply or socket outlet grasp the plug not the cord in order to avoid damaging the cable To disconnect the RZ2 1 Turn off the power switch 2 Disconnect the power cord from the power supply 3 Disconnect the power cord from t...

Страница 5: ...ted on the front panel of the RZ2 chassis contain sensitive electronics Refer to the RZ2 Technical Specifications in the System 3 Manual before making any connections The RZ2 is equipped with several different analog and digital I O capabilities The high speed and legacy fiber optic ports allow a direct connection to Z Series or Medusa Preamplifiers The RZ2 also includes onboard D A and A D conver...

Страница 6: ... connecting the RZ2 to any PZ Preamplifier ensure that the wires are inserted correctly The RED label on the back of the RZ2 denotes the input for the red colored fiber optic wire Similarly another label titled RED is shown on the back of the PZ Preamplifier ...

Страница 7: ... the plug not the cord in order to avoid damaging the cable Do not attempt to disassemble the power supply or chassis by removing any of the exterior housing of the RZ2 If you experience any issues contact TDT support immediately Do not attempt to alter this device in any way that deviates from its intended operation as described in this operator s manual Capacitors contained inside the device may...

Страница 8: ...ate pops free gently slide the fuse plate and housing out of the power supply 4 The housing contains two fuses Replace the defective broken fuse s with a new 5 x 20mm 250V 1 6A rated slow blow fuse by gently pushing the end of the fuse into the fuse housing 5 Push the fuse housing back into the power supply and press gently until the clip clicks into place Cleaning the RZ2 Processor Device To clea...

Страница 9: ... 3U Width Standard 19 rack mount Power Supply Integrated Maximum Working Voltage HI to earth ground 230V max LO to earth ground 230V max Main Voltage Rating 115 230 V 50 60 Hz 40 VA AC Installation Category CAT II Environmental Operating Temperature 0 to 40 C Storage Temperature 5 to 40 C Humidity 80 for temperatures up to 31 C decreasing linearly to 50 RH at 40 C Maximum Altitude 2 000 m Pollutio...

Отзывы: