Symmetry
When SYM is selected the SYMMETRY control varies the duty cycle from 1:19 to 19:1 to produce
sawtooth and variable pulse-width waveforms. The indicated frequency is divided by 10 with SYM
selected.
INPUTS
Sweep Input
Input Impedance:
10kOhm
Input Sensitivity:
0 to 4V for 1000:1 sweep
Maximum Allowable Input
Voltage:
±10V
Sweep Linearity:
Better than 1%
Maximum Slew Rate of
Input Voltage:
0.1V/us
Trigger Input:
Frequency Range:
DC - 5MHz.
Signal Range:
TTL compatible levels; maximum input 20V.
Minimum Pulse Width:
50ns.
Input Impedance:
Typically 2kOhm.
OUTPUTS
50 Ohm
Amplitude Control:
>20dB vernier control within each attenuator range. Maximum output
20V peak-to-peak from 50Ohm (10V into 50Ohm).
Attenuator:
Additional switch-selectable attenu- ation of 0, -20 or -40dB. Minimum
output <20mV peak-to-peak from 50Ohm (<10mV into 50Ohm).
DC Offset Control Range:
±10V from 50Ohm. DC offset plus signal peak limited to ±10V (±5V
into 50Ohm). DC offset plus waveform attenuated proportionally in -
20dB and -40dB positions.
TTL
Amplitude:
Fixed TTL level output at frequency and symmetry of main output.
Capable of driving 20 TTL loads.
GENERAL
Power Requirements
Input Voltage:
110/120V AC nominal 50/60Hz or 220/240V AC nominal 50/60Hz,
adjustable internally. The TG501 will operate safely and meet
specification within normal AC supply variations viz. 100-130V AC
and 200-260V AC respectively.
Power Consumption:
30VA Max.
Environmental Operating
Range:
+5oC to +40oC, 20% to 80% RH.
Storage Temperature Range: -40oC to +70oC.
Size:
300mm wide x 100mm high x 230mm deep
Weight:
3.4kg
3