7
Setting of data setting
NOTE
)
When setting value of input frequency filter is changed during the measurement
process, measurement is stopped. The measurement can be started again after
completion required setting parameter.
Command
:
WC41_002000 HH comparison value setting
(
HH comparison value set to 002000
)
Response
:
Response to WC41_002000
002000
Command frame
:
STX
Device No.:00
W
C
4
1
_
0
0
2
0
0
0
ETX (BCC)
02H 30H 30H 57H 43H 34H 31H 20H 30H 30H 32H 30H 30H 30H 03H
Response
:
STX
Device No.:00
Exit code
0
0
2
0
0
0 ETX (BCC)
02H 30H 30H
41H 30H 30H 32H 30H 30H 30H 03H
Memory control command
Write commend
Write the setting data into the EEPROM
Command
:
STOR
Response
:
Exit code
Command frame
:
STX
Device No.:00
S
T
O
R
ETX (BCC)
02H 30H 30H 53H 54H 4FH 52H 03H
レスポンス:
STX
Device No.:00
Exit code
ETX (BCC)
02H 30H 30H
41H
03H
Normal exit
Memory Initialization
Setting data resets to the value during the factory shipment time. But, the transmission
speed, parity, BCC switch and device number cannot brought bring to default value.
Command
:
DEFA
ULT
Response
:
Exit code
Command frame
:
STX
Device No.:00
D
E
F
A
U
L
T
ETX (BCC)
02H 30H 30H 44H 45H 46H 41H 55H 4CH 54H 03H
Response
:
STX
Device No.:00
Exit code
ETX (BCC)
02H 30H 30H
41H
03H
Normal exit