![Tsuruga Electric Corporation 3566 Скачать руководство пользователя страница 91](http://html1.mh-extra.com/html/tsuruga-electric-corporation/3566/3566_instruction-manual_1178559091.webp)
13
5.2.12
DATA?
(read-out of measurement data)
Function
Read out the measurement data.
Structure
DATA?
DATA?
: Measurement data output command
Transmission
DATA?
Response
1.
Resistance
measurement
OHM=+199.99kOHM,R-JUDGE=HI LO,VOLT=+0.1234V,V-JUDGE=FAIL
①
②
③
④
①
Resistance measurement data (data length = 11)
②
Resistance judgement output (data length = 13)
③
Voltage measurement data (data length = 13)
④
Voltage judgement output (data length = 12)
2.
Ratio
measurement
RATIO=+123.4%,Rs=1.0000̲OHM,Rx=+1.2345̲OHM,R-JUDGE=GO̲̲̲̲̲̲̲̲̲,
①
②
③
④
VOLT=+0.1234V,V-JUDGE=FAIL
⑤
⑥
①
Shows the ratio data (data length = 7)
②
Shows the resistance measurement data (data length = 13)
③
Shows the ratio standard resistance value (data length = 14)
④
Shows the resistance judgement output (data length = 13)
⑤
Shows the voltage measurement data (data length = 13)
⑥
Shows the voltage judgement output (data length = 12)
Note
: R-JUDGE
output
At
GO
:
R-JUDGE=GO
At HI
:
R-JUDGE=HI
At LO
:
R-JUDGE=LO
At HI LO
:
R-JUDGE=HI LO
At
No
judgement
output :
R-JUDGE=NULL
At
CC
error
:
R-JUDGE=CC
(CC error means the SOURCE open.)
(At the CC error, the CC error is output regardless of the judgement result.)
V-JUDGE output
At good voltage judgement
:
V-JUDGE=PASS
At
faulty
voltage
judgement
:
V-JUDGE=FAIL
At
No
judgement
output :
V-JUDGE=NULL