7
3.2.4 External control input: INT./EXT.
By making “L” level, the following operations can be externally controlled.
In this case, the operation by the switches on the front panel is disabled.
●
Selection of function
●
Selection of range
Note
:
●
The external control is not possible during the hold.
●
No selection of the function or range is possible while the remote operation of
the
memory
mode.
3.2.5
Function
input:
FUNCTION
By designation of the function code, each function can be externally controlled.
When making an input of the function code, make the INT./EXT. input “L” level in
advance.
Function
8 4 2 1
Resistance measurement
H
L
L
H
Ratio
display
L H H H
Note
: The setting other than the above is invalid.
3.2.6
Range
input:
RANGE
By making an input of a range code, the resistance range can be externally selected.
When making an input of the range, make the INT./EXT
.
input “L” level in advance.
RANGE
4 2 1
Range
L L L
Resistance
range
Auto-range
L
L
H Resistance range 30m
Ω
L
H
L
Resistance range 300m
Ω
L
H
H Resistance range 3
Ω
H
L
L
Resistance range 30
Ω
H
L
H Resistance range 300
Ω
H
H
L
Resistance range 3k
Ω
○
Timing chart of the changeover of range
1
RANGE input
STROBE
INT.EXT.
Changeover of
2
3
1
2
3
internal range