4

Calibrate the SoundPro (Pre/Post Cals)

1.  Ensure the SoundPro is turned On and is either 

stopped or paused.

2.   Attach the calibrator and cal adapter to the 

SoundPro. Set the calibrator to 1 KHz and 114 dB  
(if it is a selectable).

3.  From the Start screen, press the 

Cal

 softkey and 

the Cal screen will open. Press         when Calibrate is 
highlighted. The Pre-Cal screen will open.

4.  Allow the measurement to stabilize and press             

Arrows

 to set 114.0 dB level (see A). Then, press         

key. When set, the Calibration history screen 
appears (see B). 

5. Press          key to exit the Cal screen. 

6.  For a 

Post Calibration,

 ensure the SoundPro is 

in “Pause” mode.  (Essentially, run a measurement 
and press the run/pause key.)  The Pause indicator 
will appear at the top of the display.) Then repeat 
the Calibration steps above and press Enter to store 
the Post Calibration information. A sample screen is 
displayed below with post calibration history.

Microphone

Place calibrator 

over microphone 

QC10/QC20

Select 1kHz  

and 114 dB 

Cal Softkey

+  Select softkey  

to open

A. Calibrate

Press corresponding 

softkey to 114.0 dB 

level.  

B.  Pre-Cal

Displays  

pre-calibration 

information

Содержание Soundpro DL Series

Страница 1: ...SoundPro SE DL Series Sound Level Meters QUICK START GUIDE ...

Страница 2: ...n time 00 00 00 Softkey Menu Measurement type Repeatedly press 1st softkey to toggle through to select 1 1 1 3 SLM if purchased Backlight Key Stop Key Left Arrow Key Up Down Arrow Keys Run Pause Key Softkeys AltF Key Enter Key Right Arrow Key On Off and ESC Escape Key ...

Страница 3: ... Resetting 1 Format the card or delete the session files from the SoundPro optional To format select the File softkey from the start screen Press the down arrow continuously to select Format Press key Settings 1 Adjust the settings to fit the specifications of your sound study To open the Setup menu from the start screen press the arrow keys and press key when selected The setup menu will appear T...

Страница 4: ...ess key When set the Calibration history screen appears see B 5 Press key to exit the Cal screen 6 For a Post Calibration ensure the SoundPro is in Pause mode Essentially run a measurement and press the run pause key The Pause indicator will appear at the top of the display Then repeat the Calibration steps above and press Enter to store the Post Calibration information A sample screen is displaye...

Страница 5: ...rocess until your measurements are complete by press the Stop key This will end the session 1 From the Start screen select either View Session or select View Current Study by press the arrows and key to select 2 Press key to begin measuring Press key to Pause optional At any time press to end the session study Run Time Measurement Value Average SPL Lookup Softkey Select to change measurement param...

Страница 6: ...e bar appears and usually fluctuates if the measured value is above the minimum value for the selected measurement range Community Noise Screen This screen shows the results of measurements commonly associated with community noise studies Dosimetry Screen The Dosimetry screen provides several types of measurements of the accumulated personal exposure to noise Amplitude in the Band Shows SPL averag...

Страница 7: ...lick the button See B C Once downloaded click the Data Finder tab if you did not select go to session finder in step A See C 4 Selecting and Viewing Data A The Data Finder page is used to view select and preview reports Note Click Browse to view all data files B Select the data by clicking on either the session study Note you may need to expand to see the study data C To view in Graphs and Tables ...

Страница 8: ...t our website www tsi com for more information North America and Asia Pacific Telephone 1 800 680 1220 USA 1 651 490 2860 Outside USA Fax 1 651 490 3824 E mail technical services tsi com Europe Middle East and Africa Telephone 49 241 52303 0 Fax 49 241 52303 49 E mail tsigmbh tsi com P N 053 776 Rev B 2018 TSI Incorporated ...

Отзывы: