Troubleshooting
35
Symptom
Possible Cause
Corrective Action
Zero count above
1.2 particles/cc
Leak in the flow system.
______________________
Leak in water trap.
______________________
Restricted flow.
______________________
Pump failure.
______________________
N95-Companion
interconnect cable failure.
If you recently cleaned the
particle classifier, verify
that the O-ring is in place
and the tubing is connected
correctly.
______________________
Tighten screw and fittings.
______________________
Check the internal and
external tubing for kinks or
other blockage.
______________________
Return the instrument to
TSI for servicing.
______________________
Make sure the green light
comes on and the pump
turns on when you plug the
N95-Companion into the
P
ORTA
C
OUNT
Plus. If not,
replace the N95-
Companion interconnect
cable.
Very high zero
count (does not
zero when zero
count switch is
pressed)
Water has been drawn into
N95-Companion
.
Use water trap.
______________________
Replace sponge with dry
one.
______________________
Dry out N95-Companion
a)
Leave all components
connected including
water trap and
N95-Companion
interconnect tube.
b)
Let P
ORTA
C
OUNT
and
N95-Companion
run
overnight to dry out unit
Repeat zero count
procedure.