Chapter 4
28
The transparent cover on the water trap allows you to visually
inspect the condition of the sponge and the trap's interior. If you
wish to clean the interior or clean/replace the sponge:
1.
Remove the two screws that hold the cover in place and
remove the cover, along with the gasket.
2.
Clean the trap and sponge with alcohol to sterilize them.
3.
Dry the trap and sponge thoroughly before reassembling them
to prevent the growth of molds.
One extra sponge is included with the N95-Companion. You can
order additional sponges from TSI if necessary.
Particle Classifier
The particle classifier requires annual cleaning under normal usage
conditions. TSI recommends that the N95-Companion receive
factory service each year. It is best (but not a requirement) to send
the N95-Companion and the P
ORTA
C
OUNT
Plus to TSI at the same
time so the two instruments can be checked out together. A problem
that appears to be caused by the N95-Companion may actually be
caused by the P
ORTA
C
OUNT
Plus.
The cleaning procedure that follows should only be attempted in the
event of an instrument malfunction. It is not a substitute for the
recommended annual factory service.
Low particle counts can be caused by a dirty particle classifier;
however, there are several other possible causes that are much more
likely and should be investigated first. Consult the chapter on
troubleshooting for details.