10 CALIBRATION CERTIFICATE
Certificate of Calibration
Issued by:
TSI
Flowmeters
Ltd
24
Rockdale,
Mountrath
Road,
Portlaoise,
Co.
Laois,
Ireland.
Tel:
087-235-2107
Email:
The readings obtained by the Test Equipment referred to on this certificate are
traceable to National and International standards. Calibration Cert No. 6825CN.
Cert No.
Date of Issue
Order No.
Job No.
-01
Company Name:
Details of Flowmeter Calibrated
Type of Meter:
TSI Model EMF-300A Magflow
Serial No.
Date of Calibration:
Method of Test:
Comparison with Master Flowmeter.
The zero offset was adjusted to read 0 lpm. Since this is the first calibration of a new
flow meter there are no before service readings The unit was calibrated at a flow rate of
500 lpm. This flow rate was maintained for five minutes and correct reading was
observed on the unit. Totaliser readings were taken at one minute intervals; the maximum
deviation observed was within the specified ±2% of master flow meter readings. Correct
flow rate was verified for this calibration setting at 300, 400, 600 and 700 lpm.
Calibrated
by:
TSI Flowmeters Ltd
Version 2.0 (February 2012)
TSI Flowmeters Ltd.
25