Analog
setup screen sets the parameters that will drive the analog out port.
Applies to the 8530 Desktop model only.
Analog out [On, Off]
Turns analog out port on.
Size Fraction
Selects the size channel that will drive the
analog out.
Output Setting [V, mA]
Select between 0 to 5 V and 4 to 20 mA.
Lower Limit [mg/m
3
]
Mass concentration reading of the selected
channel that will correspond to 0 V or 4 mA.
Upper Limit [mg/m
3
]
Mass concentration reading of the selected
channel that will correspond to 5 V or 20 mA.
Settings
Settings
screen sets basic unit parameters.
Date Time
Sets current date, current time and date/time format.
Time can be set in 12 or 24 hour format. Date can be
set in yyyy/dd/mm, yyyy/mm/dd or yyyy/dd/mm.