TSI DustTrak DRX Desktop SOP v1.1

5 |

P a g e

b. For the 2

nd

format go to File -> Export -> Export Test Data… In the new window select the test

with the mouse and then select “Export”. In the new window go to the appropriate folder and

name the file based on the present day in the format MMDDYYYY

10. Once the file has been saved we then need to delete the file from the device. This is done through File ->

Receive selecting the test with the mouse and then clicking on “Delete”. Two confirmation windows will

appear, select “Yes” only if you have successfully downloaded the data in step 10. Once done, close the

window.

11. Now update the device clock. This can be done under Instrument Setup -> Parameters -> Clock…

a. First note the time (and date) if any difference exists between the PC and the DRX device and

mark it in the log.

b. If there is a difference update the clock by setting the “Instrument Time” to a time in the future

by 10-20 seconds (also be sure that the date is correct). And then 3 – 4 seconds ahead of

reaching that time press the “Send” button. This needs to be done as there is a delay in the

transmission. You will get a pop-up window confirming that the time has be changed and you

have to select “OK” and then select “Close” on the clock update window.

12. To restart the device we want to coordinate the time intervals from all instruments so the best way to do this

is through Instrument Setup -> Logging Setup…

a. In this new window choose the program of interest. The logging program it was last using will

automatically be selected.

b. Under Start Date and Start Time choose today’s date and a time on an even interval of the

recording interval with enough time allowed for putting the instrument back in place. For

example, for a 2 minute recording interval choose a Start Time of 14:10 or 14:12 or 14:14 or

14:16, etc.

c. Select “Send” on the right hand side once the options are correct.

13. To actually start the DRX device you still need to tell the instrument to start logging. This is done in the same

window as stopping it: Instrument Setup -> Logging Control

a. Select “Discover Instrument”

b. Once the DRX device info appears check that the right program logging program is selected and

then choose “Start Logging”. Then close the window.

14. At this point the USB cable can be disconnected and the TrakPro software closed and the instrument

returned to its proper measuring height.

2. Setup/First start:

Use this guide when you are setting up the device for the first time.

1. Make sure the DRX device is turned on and has loaded up to the home screen, the instrument is connected

via the USB Device port on the DRX device to one of the USB ports on the PC, and the TrakPro software has

been loaded.

2. At this point you will need to wait until the proper USB drivers are loaded (you may or may not see this

activity in the start bar near the PC clock).

3. Once they have loaded, you can initiate communications between the device and the PC by going to

Instrument Setup -> Communications…

4. A window will pop up labelled “NDIS Communications” and should give details of the Model, Serial Number,

Status, and IP address of the connected DRX device.

a. If the above happens go to the next step

b. If the above does not happen, wait a couple seconds and click the “Test” button.

c. If the option in b. is still not working the best option is to restart the DRX device using the ON/OFF

button and try again from the beginning.

5. Once you have an open line of communication we first need to set the date and time. This can be done

under Instrument Setup -> Parameters -> Clock…

a. First note the time (and date if any) difference between the PC and the DRX device and mark it in the

log

Содержание DustTrak DRX

Страница 1: ...k Export Name file based on the date as YYYYMMDD 17 Now delete the test files by going to File Receive select Delete All 18 Correct the time of the DRX through Setup Parameters Clock 19 Note the time...

Страница 2: ...nnected to the USB Device connection on the device Note the only option to download using a USB drive is download all and only in a comma separated file format The DRX device communicates through NDIS...

Страница 3: ...n Settings are more easily modified through a PC using the TrakPro software which will be covered in the next section 5 Go into the Zero Cal option be selecting it This is where one can manually zero...

Страница 4: ...of the connected DRX device c If the above happens go to back into the Logging Control following step 3 d If the above does not happen wait a couple seconds and click the Test button e If the option i...

Страница 5: ...a 2 minute recording interval choose a Start Time of 14 10 or 14 12 or 14 14 or 14 16 etc c Select Send on the right hand side once the options are correct 13 To actually start the DRX device you stil...

Страница 6: ...fore it starts to record iii Log Interval This is how often the data is recorded in the final memory For a 2 minute data interval required this should be set to 2 minutes iv Test Length This is how lo...

Страница 7: ...sten to the pump while running the DRX device in Manual RunMode Under Setup FlowCal on the DRX unit interface the speed of the pump can be increased decreased make sure there is an audible coordinatin...

Отзывы: