HEATED
INLET
SAMPLE
CONDITIONER
ACCESSORY
(854041)
FOR
DUSTTRAK
™
ENVIRONMENTAL
MONITORS
INSTALLATION GUIDE
Страница 1: ...___ TSI and TSI logo are registered trademarks of TSI Incorporated DustTrak is a trademark of TSI Incorporated HEATED INLET SAMPLE CONDITIONER ACCESSORY 854041 FOR DUSTTRAK ENVIRONMENTAL MONITORS INSTALLATION GUIDE ...
Страница 2: ......
Страница 3: ...owing table are present 2 Contact TSI immediately if items are missing or broken 3 Additional items may be included if you ordered accessories or spare parts Qty Item Description Reference Picture 1 Heated Inlet Column 1 RH Temp Sensor 1 RH Temp Sensor Gland 1 Heated Inlet Control Module 1 Omni Direction Inlet Qty Item Description Reference Picture 1 Water Trap 1 Power and Communication Cable 1 Co...
Страница 4: ... ensure a waterproof seal 3 Insert RH Temp Sensor through gland and tighten gland nut Extend probe 2 75 7 cm beyond nut 4 Attach the sun shield to the top of the RH Temp Sensor as shown Sun Shield should not touch sintered cap of RH Temp Probe 5 Attach RH Temp Sensor cable to heated inlet control module J3 6 Attach the two larger gray connectors to side of the photometer 7 Attach the two smaller w...
Страница 5: ...umn 12 Attach the inlet to the column 13 Install O ring and route cable through the mounting ring Pull excess cable through to prevent pinching of cable when inlet column is installed 14 Set inlet column on top of photometer mounting ring rotate slowly until alignment pin mates with the hole in the bottom of heated inlet column 15 Slide column retainer ring down heated inlet column 16 Hand tighten...
Страница 6: ...nscrew the top knurled ring collar from the lower mounting collar 2 Inside the Enclosure disconnect the Heated Inlet cable note that the connector has a lock release on it 3 Remove the Ferrite Clamp Assembly Note that the clamp is released by inserting and twisting a small flat blade screwdriver Also for purposes of re assembly note that the cable twists two times through the Ferrite 4 Carefully p...
Страница 7: ...escribes the status of the heated inlet Solid Green Inlet temp is 1 C below set point and controlling to maintain set point Blinking Green Inlet temp is between 1 to 5 C below set point and controlling to improve Blinking Red Inlet temp is more than 5 C below set point and controlling to improve This will occur when unit is first turned on and coming to temperature Solid Red Sensor unplugged or ha...