16
16
6. Enter the filename under which you will store
the data.
7. Choose
the
OK
key.
Serial Communication Through
Windows 95
Click on
Hyperterminal.exe
located in the
accessories menu.
If a box comes up requesting modem
installation, click
No
.
Enter a name (Alnor) and choose an icon for
meter downloading to the PC, click
OK
.
Connect using
Com1
connection, click
OK
.
Set port settings under
Com1 Properties
to
1200BPS, 8 data bits, no parity, stop bits 1,
and
Flow Control
to
None
, click
OK
.
Click
File
and choose
Properties
.
Click on
Settings
, choose
ASCII
set up.
Verify checkmark for
Terminal Keys
,
Emulation
to
Auto Detect
, and
Backscroll Buffer Lines
to
500
.
Place a check mark under
ASCII Receiving
in the boxes marked
Append Line Feeds
To Incoming Line Ends
and
Wrap Lines
That Exceed Terminal Width
, click
OK
.
Click
O
K
again to exit
Properties
.
Hyperterminal
is now ready to receive
information from the meter.
Save terminal configuration for future use.
Appendix D:
Pressure to Velocity Conversion Charts
The AXD 550 MicroManometer converts pressure
readings to velocity using the formula for standard
conditions:
V
(fpm)
= 4005
P
∆
Pressure in inches H
2
0
V
(m/s)
= 1.29
P
∆
Pressure in Pa
Please note that for small pressure readings, an
increment in pressure can cause large steps in
velocity, but at larger readings, the steps in
velocity are much less pronounced as a percentage
of reading. A Micro-Manometer should not be
used with a pitot tube at very low differential
pressures to measure velocity, because resolution
errors will be large.
The following graphs show approximate velocities
for certain differential pressures in the range0–
0.10 inches H
2
0 (0–25 Pa).