11
MODEL RVA801/LCA301
SPECIFICATIONS
(subject to change without notice)
Measurement Range
Velocity
50 to 6000 fpm
0.25 to 30 m/s
Temperature 32 to 140
°F
0 to 60
°C
Volumetric
Flow Rate
1-9999000 cfm
1 - 999999 l/s
1 - 999999 m
3
/h
0.01 – 3000 m
3
/s
Area Input
0.043 - 900 ft
2
0.00399 - 90 m
2
Actual range is a function of velocity and area input.
Display Resolution
Velocity
1 fpm
0.01 m/s
Temperature
1
°F
0.1
°C
Accuracy
Velocity
±1% reading ±4 fpm
±1% reading ±0.02
m/s
Temperature
±2 °F
±1 °C
Storage
Temperature
14 to 140
°F
-10 to 60
°C
Operating
Temperature
32 to 140
°F
0 to 60
°C
Power Source
9-volt battery
Battery Life
Approximately 40 hours continuous use
Dimensions
(LWD)
11 in.
×
4.5 in.
×
2.6 in.
280 mm
×
112 mm
×
65 mm
Weight
10 oz.
275 g
(excluding battery)