44
Appendix B
Type
Value
Definition
Actions/Resolutions
Device
Status
(cont.)
16
Device internal time
has not been synced
to the cloud in the
last 24 hours
If error persists may have
to replace entire sensor
cartridge, it is not possible
to replace the coin battery
on the backside of the
sensor board
32
Device internal time
value is invalid due to
low battery,
unpowered device
and lack of cloud
connection
If error persists may have
to replace entire sensor
cartridge, it is not possible
to replace the coin battery
on the backside of the
sensor board
64
Sensor board
EEPROM failure.
Some sensors may
be disabled.
Replace sensor board
Sensor
Status
Messages
(for all
sensor
types, such
as CO
2
,
NO
2
, O
3
,
CH
2
O,
Temp/RH,
PM, etc.)
0
No errors
N/A
1
Sensor is not factory
calibrated
Replace sensor with
factory calibrated sensor
2
Sensor hardware
fault
Power cycle device.
Disconnect power and
check sensor connections
to ensure sensor cables
are in good working
condition and sensors are
plugged in. If error persists,
replace individual sensor
or complete sensor
cartridge
4
Sensor
communication error
Power cycle device.
Disconnect power and
check sensor connections
to ensure sensor cables
are in good working
condition and sensors are
plugged in. If error persists,
replace individual sensor
or complete sensor
cartridge
Содержание AIRASSURE 8144
Страница 6: ...iv Appendix A Specifications 39 Appendix B Status Flags and Error Codes 43 Examples 46 ...
Страница 35: ...Operation 25 ...
Страница 36: ...26 Chapter 3 This page intentionally left blank ...
Страница 44: ...34 Chapter 4 This page intentionally left blank ...
Страница 48: ...38 Chapter 5 This page intentionally left blank ...
Страница 52: ...42 Appendix A This page intentionally left blank ...
Страница 57: ......