3-16
AeroTrak
®
Handheld Airborne Particle Counter
5. Press the
Timing
tab to enter or edit start delay times, sample time,
hold time, etc.
6. To make changes to the timing settings, highlight the value to change
(hours, minutes, seconds, etc.) and use the on-screen keypad to
change the value.
W A R N I N G
Instrument status alarms are inactive if samples times are 10 seconds
or less. Flow error alarms may not occur is sample time is less than
10 seconds. To ensure proper instrument status and flow alarm
operation, use a sample time of 15 seconds or larger.
NOT I CE
With firmware version 2.4 and higher, you may select a date to print
reports and/or export sample data. There is no need to create new
zones to select data by date. There is no need to print all samples in a
zone to view a specific sample or samples taken on specific dates.
Samples can also be selected by date for a given recipe and zone.