Installation and Getting Started
2
–5
Ins ta llat ion
Installation of the A Remote Particle Counter consists of:
•
Determining the installation location
•
•
Supplying power to the particle counter
•
Connecting communications from the particle counter to the
computer
•
Connecting the vacuum and sample tubes to the particle counter
Determine the Installation Location
Determine the installation location according to your monitoring needs.
Remote particle counters are designed so that they can be located close to
critical locations determined by risk analysis. The Remote Particle Counter
should be mounted such that the sensor assembly can be easily installed
and removed for calibration. It is generally convenient to mount the particle
counter to a vertical flat surface such as a wall, but the particle counter can
also be mounted below the critical work area, inside a vented hood, or a
convenient location close to the point of measurement.
Mounting the Remote Particle Counter
The Remote Particle Counter can
be mounted using a variety of
mounting brackets and schemes.
TSI provides an optional mounting
bracket (TSI P/N 700008 for 0.1 cfm
models and 700008-1 for 1 cfm
models) that allows the particle
counter to be easily mounted and
removed on a surface (see figures).
To install the mounting bracket:
1. The bracket is provided with two
0.169-inch (4.30 mm) diameter
holes, suitable for a #6 or M4
screw. Screw the bracket to the
mounting surface using
appropriate screws.
2. Slide the particle counter onto
the mounting tabs at the top of
the bracket and snap it into the
locking tab (circled in figure).
To remove the Remote Particle
Counter from the mounting bracket:
1. Press the locking tab at the top
of the bracket.
2. Lift the particle counter off the
mounting tabs.
Mounting Bracket
(optional)
Remote Particle Counter Mounted
on Optional Bracket
Содержание AEROTRAK+ 6201
Страница 2: ......
Страница 12: ...x AeroTrak Remote Particle Counter Operation Manual This page intentionally left blank ...
Страница 39: ...Communications 3 13 Non Pump Models Pump Models ...
Страница 46: ...4 2 AeroTrak Remote Particle Counter Operation Manual This page intentionally left blank ...
Страница 52: ...A 4 AeroTrak Remote Particle Counter Operation Manual Dimensional Diagram Models 6201 6301 6501 ...
Страница 55: ......