4
Chapter 2
Extending the Probe
To extend the probe, hold the handle in one hand while pulling on the
probe tip with the other hand. Do not hold the cable while extending the
probe as this prevents the probe from moving.
Retracting the Probe
To retract the probe, hold the handle in one hand while pushing on the
probe tip with the other hand. If you feel the probe antenna binding, pull
gently on the probe cable until the smallest antenna section is retracted.
Collapse the rest of the antenna by pressing the probe tip.
Articulating Probe (Models 8384A/8385A/8386A Only)
The articulating probe has the ability to bend at a 90° angle for those
hard to reach places. To bend the probe, loosen the knurled nut on the
joint, bend the probe, and tighten the nut. To straighten the probe,
loosen the nut, straighten the probe, and tighten the nut.
Changing the Real-Time Clock
The V
ELOCI
C
ALC
Plus has an internal real-time clock that keeps track of the
time of day (the format is HH.MM where HH is the hour in 24-hour format
and MM is minutes) and the date. It is very important to set the time and
date correctly, otherwise date and time stamping of recorded data will not be
correct. This information has been set to Central Time at the factory before
shipping.
To change the time and date, press and hold both
and
keys during the
power-up sequence when the time is displayed. Release the keys when the
V
ELOCI
C
ALC
Plus beeps twice. You will have an opportunity to view and/or
change the hours, minutes, year, month, and day of month in sequence. Use
the up and down arrow keys (
) to change any settings. Use the
ENTER
key to store each setting and advance to the next one.
Changing the Baud Rate
The V
ELOCI
C
ALC
Plus has a variable baud rate that is used when
downloading or printing data from the instrument. By changing the baud rate
to a higher rate, the data will be downloaded faster.
NOTE: The baud rate
must be equal to that of your computer or printer.
The instrument baud rate is displayed during the initial power up sequence.
To change the baud rate, press and hold both
and
keys during the
power-up sequence while the baud rate is displayed. Release the keys when
the V
ELOCI
C
ALC
Plus beeps twice. Use the
and
keys to scroll through
the available values of 1200, 2400, 4800, 9600, and 19200. Press
ENTER
to
set the value that is displayed.