![TSI Incorporated VELOCICALC Plus 8360 Скачать руководство пользователя страница 21](http://html1.mh-extra.com/html/tsi-incorporated/velocicalc-plus-8360/velocicalc-plus-8360_operation-and-service-manual_1178089021.webp)
15
Appendix A
Specifications
Model 8360/8388
V
ELOCI
C
ALC
Plus
Models 8382/8383
V
ELOCI
C
ALC
Plus
Velocity
Measurement Range
Accuracy
¹
1
30 to 9,999 ft/min
(0.15 to 50 m/s)
3% of reading or 3 ft/min (0.02
m/s), whichever is greater
Same
Same
Temperature
Measurement Range
Resolution
Accuracy
14 to 140°F (-10 to 60°C)
0.1°F (0.1°C)
±0.5°F (0.3°C)2
0 to 200°F (-17.8 to 93.3°C)
Same
Same
Instrument Temp. Range
Electronics
Operation
Storage
Probe
Operation
Storage
40 to 113°F (5 to 45°C)
-4 to 140°F (-20 to 60°C)
14 to 140°F (-10 to 60°C)
-4 to 140°F (-20 to 60°C)
Same
-22 to 194°F (-30 to 90°C)
0 to 200°F (17.8 to 93.3°C)
-22 to 194°F (-30 to 90°C)
Relative Humidity Range
Accuracy
0 to 95%
±3% rh3
N/A
N/A
Dew point Range
5 to 120°F (-15 to 49°C)
N/A
Static/Differential
Pressure
Range4
Accuracy5
-10.00 to +10.00 inches H
2
O
(-2.500 to +2.500 kPa or -20.00 to
+20.00 mm Hg)
±0.5% of rdg ±0.01 in H
2
O
(±0.002 kPa or ±0.02 mm Hg)
Same
Same
Volumetric Flow Rate
0.2 to 4,50,000 ft3/min (0.0424 to
1,170,000 m3/hr or 0.1 to 325,000
l/s)
Same
Duct Sizes
1 to 100 inches in increments of
0.5 inch, 100 to 255 inches in
increments of 1 inch
(1 to 100 cm in increments of 0.5
cm, 100 to 255 cm in increments
of 1 cm)
Same
Averaging Capability
Up to 255 values each of velocity,
temperature, pressure, humidity,
or pressure flow
Up to 255 values each of
velocity, temperature,
pressure, or pressure flow
Содержание VELOCICALC Plus 8360