9
Chapter 4
Maintenance
The Model 9545/9545-A requires very little maintenance to keep it
performing well.
Recalibration
To maintain a high degree of accuracy in your measurements, we
recommend that you return your Model 9545/9545-A to TSI for annual
recalibration. Please contact one of TSI’s offices or your local distributor to
make service arrangements and to receive a Return Material Authorization
(RMA) number. To fill out an online RMA form, visit TSI’s website at
U.S. & International
TSI Incorporated
500 Cardigan Road
Shoreview MN 55126-3996
Tel: (800) 874-2811
(651)
490-2811
Fax: (651) 490-3824
The Model 9545/9545-A can also be recalibrated in the field using the
CALIBRATION menu. These field adjustments are intended to make minor
changes in calibration to match a user's calibration standards. The field
adjustment is NOT intended as a complete calibration capability. For
complete, multiple-point calibration and certification, the instrument must be
returned to the factory.
Cases
If the instrument case or storage case needs cleaning, wipe it off with a soft
cloth and isopropyl alcohol or a mild detergent. Never immerse the Model
9545/9545-A. If the enclosure of the Model 9545/9545-A or the AC adapter
becomes broken, it must be replaced immediately to prevent access to
hazardous voltage.
Storage
Remove the batteries when storing the unit for more than one month to
prevent damage due to battery leakage.