Standard Velocity vs. Actual Velocity
33
Example No. 1:
You want to measure the actual velocity in a duct.
The air temperature in the duct is 55°F and the
pressure is 14.24 psia. You take a measurement
and the display reads 1200 feet per minute (ft/min).
ActualVelocity
=
+
+
=
1200
460 55
460 70
14 7
14 24
1203 7
.
.
.
ft / min
Example No. 2:
You need to measure the actual velocity in a
plenum. The air pressure is 99.4 kPa and the
temperature is 27°C. The display reading is 2.3
meters per second (m/s).
ActualVelocity
2.3
=
+
+
=
273 27
273 211
101
99
2 39
.
.4
.4
.
/
m s