Symptom
Possible
Causes
Corrective Action
Display
reads "LO"
Low battery
charge
Replace or
recharge batteries.
Wrong
AC
adapter
Replace with the
correct AC
adapter.
Low AC line
voltage
Correct the AC
line voltage or use
batteries.
Dirty
battery
contacts
Clean the battery
contacts.
Temperature
initially
reads high
Temperature
sensor is still
warm from
velocity
mode
Allow about 30
seconds before
reading
temperature.
Display
reads "ERR"
You are
trying to
enter more
than 255
readings
Read or record the
average; clear the
sample register
and proceed.