CaloSys Modules
55
Information about Reference Air
11.4.2.
The reference air is one of the crucial factors for the exact determination of the O2 and CO2
concentrations during the measurement. Rapid alterations in the reference air will lead to incorrect
results. For this reason, you should ensure that…
•
…as far as possible, nobody should enter the test area while a measurement is being made,
no doors or windows should be open in the measuring area, and that there should be no
drafts.
•
…no short-term variations in the room climate can occur.
•
…the CaloSys module is calibrated under the same conditions as used in the measurement.
Remember that daily or seasonal variation may also affect the room climate.
11.5. Connecting PC
Connections
Ports
PC
Calorimetry Master Control Unit
"X01", "X02" - PCI-9111
"X01", "X02"
"X03" - PCI-9114
"X03"
Fig. 29 PC ports
System configuration
This section describes the connections of a PC of a CaloSys module to be operated with manual
calibration (see Sections 11.1.1.1 (push principle), 11.1.2.1 (pull principle)). For instructions about
connections of the PC of CaloSys modules to be operated with manual or automatic calibration, see
Section 11.1.1.2 (push principle), 11.1.2.2 (pull principle). For instructions about connections of the
PC of CaloSys modules retrofitted for automatic calibration, see Section 11.3.1.
Three cable connections are used to connect the PC to the Calorimetry Master Control Unit -
connections with the same markings are connected to each other as shown in the table.
11.6. Air Supply Pump
System configuration
This section describes air supply pumps of CaloSys modules working with push principle (see Section
11.1.1). For instructions of vacuum pumps of CaloSys modules working with pull principle (see
Section 11.1.2), see Section 11.8.
NOTICE
- Reduced lifetime!
If Calorimetry Process Control Units and Air Pumps are powered ON without operating the
system, this may reduce lifetime of devices.
•
Always run system software when devices are powered ON.
•
Power OFF devices after use.
The air supply pump allows to generate and maintain the set airflow within the CaloSys module.
Содержание Calorimetry PhenoMaster
Страница 6: ...6 PhenoMaster Hardware Operating Instructions info TSE Systems com www TSE Systems com ...
Страница 13: ...CaloSys Modules 13 Confirm default installation path by pressing Next info TSE Systems com www TSE Systems com ...
Страница 15: ...CaloSys Modules 15 Press Finish info TSE Systems com www TSE Systems com ...
Страница 69: ...CaloSys Modules 69 info TSE Systems com www TSE Systems com ...